Specifications

Operating Manual SEFRAM 4451
M4451 A00
27
3.6.4 MODE Menu
Selects the output mode: CONT (Continuous), TRIG (Triggered), GATE (Gated), BRST (Burst) and
PHASE.
To select the output mode, press MODE, then press the function key that corresponds to the desired Mode
menu option, as shown:
Mode Menu
F1: CONT - (Continuous) - Selects continuous output.
F2: TRIG - (Triggered) - Triggers one output cycle of the selected pulse for each trigger event.
F3: GATE - (Gated) - Triggers output cycles as long as the trigger source asserts the gate signal.
F4: BRST - (Burst) - Triggers output N output cycles for each trigger event, where N ranges from 2 to
999,999.
F5: EXTWID - In the external width (EXT WID) pulse mode, the pulse period and width are determined
by the externally applied signal. The SEFRAM 4451 then applies transition and level
parameters to this signal in order to generate the pulse. The period, width and delay may be
changed, but their change has no effect on the pulse, and their values are not displayed. The
trigger mode may not be changed while the external width pulse mode is enabled.
External Pulse
After selecting the TRIG , GATE or BURST menu, the trigger source menu is available: