Specifications
Operating Manual SEFRAM 4451
M4451 A00
26
3.6.3 PULSE Menu
PULSE Menu
F1: RISE - Selects the pulse Rise time (Leading edge).
F2: FALL - Selects the pulse Fall time (Trailing edge).
F3: EQUAL - Selects equal Rise (Leading edge) and Fall (Trailing edge) times.
F5: NORMAL/COMPL - Selects the Normal or Complement pulse mode.
COMPLEMENT Pulse Mode
The transition time range is 5ns to 100ms, but the value is limited to a 20:1 ratio between the
transition times. In addition, both values must be within one of the following ranges:
5ns - 100ns
50ns - 1us
500ns - 10us
5us - 100us
50us - 1ms
500us - 10ms
5ms – 100ms
The transition times are also limited by the values of the period, width and delay.