User's Manual
39
(30% dip in
UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 seconds
continued operation
during power mains
interruptions, it is
recommended that
the mobiCARE-
MC200M series be
powered from an
uninterruptible
power supplied or a
battery
Power
frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment
Note UT is the a.c. mains voltage prior to the application of the test level
Guidance and manufacturer’ declaration – electromagnetic immunity
The mobiCARE-MC200M series is intended for use in the electromagnetic
environment specified below. The customer or the user of the mobiCARE-
MC200M series should assure that it is used in such an environment
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic
environment-guidance
Conductive
RF
IEC61000-4-6
3Vrms
150kHz
to 80MHz
-
10V/m
Portable and mobile RF
communications
equipment should be no
closer to any part of the
mobiCARE-MC200M series
Including cables, than the
recommended distances
calculated from the
equation applicable to the