Guidance and manufacturer’ declaration – electromagnetic immunity
The mobiCARE-MC200M series is intended for use in the electromagnetic
environment specified below. The customer or the user of the mobiCARE-
MC200M series assure that it is used in such an environment
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic
environment-
guidance
Electrostatic
discharge
(ESD)
IEC 610
00-4-2
±8 kV
contact
±2, 4, 8, 15
kV air
±8 kV contact
±2, 4, 8, 15
kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
-
Mains power quality
should be that of a
typical commercial
or hospital
environment
Surge
IEC61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
-
Mains power quality
should be that of a
typical commercial
or hospital
environment
Voltage dips,
short
interruptions
and voltage
variations on
power supply
lines
IEC6100-4-11
<5% UT
(>95%dip in
UT)
for 0.5 cycle
40% UT
(60% dip in
UT)
for 5 cycles
70% UT
-
Mains power quality
should be that of a
typical commercial
or hospital
environment
If the user of the
mobiCARE-
mobiCARE-MC200M
series requires