User's Manual

37
8. Declaration of Conformity
8.1. Electromagnetic Compatibility
It has been independently tested and it manufactured in compliance
with the following EMC standard IEC 60601-1-2:2014
Guidance and manufacturer’ declaration – electromagnetic emissions
The mobiCARE-MC200M series is intended for use in the electromagnetic
environment specified below. The customer or the user of the
mobiCARE-MC200M series should assure that it is used in such an
environment
Emissions
test
Compliance
Electromagnetic environment-
guidance
RF emissions
CISPR 11
Group 1
The mobiCARE-MC200M series use
RF energy only for its internal
function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The mobiCARE-MC200M series must
be used only in a shielded location
with a minimum RF shielding
effectiveness and, for each cable that
exists the shielded location, a
minimum RF filter attenuation of
13dB from 30MHz to 1000MHz
The mobiCARE-MC200M series, when
installed in such a shielded location,
is suitable for use in all
establishments other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
-
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
-