User's Manual

28
The mobiCARE-MC200M series is connect to the Android & iOS
Application. The data indicating the status of the device is transmitted
to the application from the device.
The required specification of the interface
- Main Device / Mobile App interface
- Data format : Bluetooth
- Mobile Application : Application on based Android & iOS
- Device status information
- Device information (Bringing information when connecting to BLE)
- Battery information (Bringing information when connecting to BLE)
- Data to Device External memory
- Transmits the data to server
4.5. Error Message & Troubleshooting
4.5.1. Error Message & LED Display
Message & LED
Description
Solution
Check the status
of the device
Can’t connect to a
patch
Check the patch connection
status .
Check the battery level and
replace it if it is insufficient.
Check the signal strength and
replace the device if the error
continues.
Battery status
Battery remaining
less than 100%
Replace the battery.
Patch signal
status
No ECG Signal
Check if the electrode is properly
connected and make sure that the
attachment is correct.
Orange LED
lights up during
measurement
External memory
error
Check if the external memory is
properly installed in the product
socket and the installation status
is correct.
Orange LED
blinks during
measurement
No ECG Signal
Check if the electrode is properly
connected and make sure that the
attachment is correct.
Green LED
Bluetooth
If the distance between the