User's Manual
27
Software
Device Firmware
mobiCARE Firmware (v1.0.0)
App software
mobiCARE Cardio Hospital (v1.0.0)
mobiCARE Cardio Diary (v1.0.0)
Web software
mobiCARE Web Service (v1.0.0)
Minimum
specification of
Smart device
OS
Android 7.0 (over)
iOS 12.0 (over)
RAM
2GB
Inside Memory
16GB
Component
mobiCAREC-MC200M / mobiCAREC-MC200ML /
mobiCARE-MC200M7 / mobiCARE-MC200ML7,
Coin Battery 1EA, Manual, Open Tool
Operation
Storage/Transport
Operating
Environment
Temperature
5 ~ 40℃
-25 °C to + 5 °C , and
+5 °C to + 35°C at a
relative humidity up to
90%, non-condensing;
>35°C to 70°C at a
water vapour pressure
up to 50hPa
Humidity
10 ~ 95%R.H
(Non-condensing)
Atmospheric
pressure
700 ~ 1060
hPa
IP grade
IP44
Technical
Characteristic
Heart Rate Range
30 ~ 280bpm
Heart Rate
Accuracy
≤30 ~ ≤120 ± 2bpm
<120 ~ ≤280 ± 3bpm
RF
Frequency
2.4 GHz band
RF Output Power
Max. 0 dBm
Manufacturer/country
Seers Technology Co., Ltd / KOREA
4.3. Maintenance
For cleaning, gently wipe the mobiCARE-MC200M series with medical
grade disinfectant. DO NOT immerse the mobiCARE-MC200M series is
water resistant, but should be kept dry. This device does not have user
serviceable components.
Do not disassemble, crush, puncture, short external contacts or circuits,
dispose of in fire or water.
4.4. Specification of External Device connection