After selecting the calibration profile, the profile characteristics are shown in a popup dialog.
Once the calibration profile has been selected, press the Setup [ ] button and select
«Calibration» «OSL» to start the OSL calibration procedure.
The calibration procedure is as follows:
Enter the start frequency (default is the
lower frequency limit)
Enter the stop frequency (default is the
upper frequency limit)