User Manual

SARK-110-ULM User’s Manual
Rev 1.0 June 22
th
, 2019 - 34 - © 2019, Melchor Varela EA4FRB
Element
Step ResponseImpulse Response
Unity reflection Unity reflection
Open
Unity reflection, -180º Unity reflection, -180º
Short
Positive peak Positive level shift
Resistance
R > Z
0
Negative peak Negative level shift
Resistance
R < Z
0
Positive then negative peaks Positive peak
Inductance
Negative then positive peaks Negative peak
Capacitance
Display