The calibrated range accuracy is described in next table. Uncalibrated or out-of-range power levels are
described as “n/a”. In these ranges the device can be used but output power level may not be accurate.
RF Explorer Signal Generator Combo output power accuracy
Frequency range (MHz)
Power range (dBm)
Typical Accuracy (dB)
Worst case Accuracy (dB)
0.1 to 1
-40 to 0
±2
±5
0 to +10
n/a
n/a
1 to 10
-40 to 0
±1
±2.5
0 to +10
n/a
n/a
10 to 1950
-40 to -10
±1
±2.5
-10 to +10
±1.5
±3.5
1950 to 3600
-30 to +10
±1
±2.5
3600 to 5200
-30 to +10
±1.5
±3
5200 to 6000
-30 to 0
±2
±4
0 to +10
n/a
n/a
Future firmware upgrade will increase accuracy in some frequency and power ranges using internal
calibration table adjusted in factory.
An optional software process will be available to manually calibrate the device and adjust for amplitude
deviation. Contact us for additional details.
Absolute maximum and minimum power levels available are described in next table. Note these levels are
not guaranteed and not calibrated.
RF Explorer Signal Generator Combo Absolute output power range
Frequency range (MHz)
Minimum output power (dBm)
Maximum output power (dBm)
0.1 to 1
-60
0
1 to 10
-60
+5
10 to 3200
-60
+15
3200 to 3650
-60
+10
3600 to 5200
-60
+15
5200 to 6000
-60
0