Manual

SEED HAWK
®
2014 30 SERIES OPERATOR’S MANUAL – DOMESTIC46
5.4.4 Ground Speed Calibration
All new systems require a ground speed calibration to ensure accurate area
totals and accurate ground speed readings.
During the calibration the monitor is counting the number of pulses produced
from the drill mounted sensor. The monitor will display the pulses counted in
400 ft. (122m). Choosing this item will allow you to calibrate the ground speed
sensor to a known distance. The default distance is 400 ft. (122m) but this can
be set by the user. To perform a ground speed calibration, determine the dis-
tance to be used and measure out a course.
1. To complete the calibration, mea-
sure a course 400 ft. (122m) long
preferably on level ground with a start
and nish point.  Line up tractor at 
the Start Point. The drill must be in
the down position throughout the
procedure. On the monitor, verify that
the calibration length is 400 ft. (122m).
If not, be sure to correct it. Once
the distance is selected, highlight and
select “Continue.
3. At the end of the 400 ft. (122m)
course, stop the tractor, highlight the
“Complete” button and press Enter.
2. The monitor now displays “Start
Driving” and shows a Wheel Pulses
Received window. You should see this
number increase as you drive the 400
ft. (122m) course.