User manual

Page 16 Security & Electronic Technologies GmbH
Software version V1.00
Serial number SN1234 X0xxxx Bxxxx
22. Test
Indication of the current mA measured value and the
hex value for the A/D-converter
23.
TestR
Rel1 ON
Rel1
OFF
Rel2 ON
Rel2
OFF
Relay test
24. Init
Sure:
No
Sure:
Yes
Abort
Clear all settings (factory settings)
26. Exit
Back to display mode
Error messages
Error Code
Significance
Error E1
The set value is invalid
Error E2
Measured value too small (<3mA => probe defective)
Error E3
Measured value too large for zero point calibration
(probe must not be immersed)
Error E4
Not defined
Error E5
Not defined
Error E6
Set height is too great (measured value is too small -
the probe must be immersed)
Error E7
Set volume is too great (measured value is too small -
the probe must be immersed)
Error E8
Measured value is too great (probe short-circuited,
defective, or incorrectly connected
Error E9
Measured value is 0 (no probe connected, or wire
breakage, or reversed polarity)
Error E10
Calibration error
Error E11
Warning „Tank empty‟ (tank content is actually too