Instructions

Version 04-2017
Setup and Operation
P. 9
Special parameters
Additional
input functions:
Description
For which
device
Menu items 1 to 7
The menu items 1 to 7 contain the basic setup of the devices.
Some special settings like language or network parameters
or others have to be set up via menu items 9 to 24.
all
9. Offset probe
Sub-menu a. Offset calibration‘ (electrical zero point)
b. ‘Probe bottom gap’ (position over ground)
c. ‘Bottom dead stock’ (shall not be displayed)
- ESC Exit this sub-menu.
- Offset calibration: Stores signal value of probe’s zero point.
Probe must not be plunged.
- Probe bottom gap: Distance: x cm
Standard is x = 0 cm, max = 99 cm
- Bottom dead stock: Sucking position over ground: y cm
Standard is 0 cm, means total content.
y > 0 cm means dead stock height which
doesn’t occur in the liter displaying
- Default values: Resets all values back to standard 0.
all
10. Trim height
Input option for the reference height for calibration of
probe and measurement device.
It is useful in case of unknown specific weight of the fluid.
Enter the beared liquid level: xx.x cm ( + / - / Enter ).
Confirm with ’Calibrate: Yes’
If this is done at a low tank filling level it is recommend to
repeat this later again at a higher filling level. Also refer to 11.
all
11. Exit
Jump further with [+] or Exit with [Enter]
all
12. Unit
Selectable units are: L (Liters), %, m, kg, t (Tons),
IG (Imp.gallons), UG (US gallons). ( + / - / Enter ).
Displaying mbar or kPa is also selectable.
‘%’ or ‘m’ values are displayed with two decimal numbers.
all
13. Rounding
Automatically - Default setting for autom. rounding
Without rounding - No rounding means highest resolution.
Maybe wobbling values. A certain
rounding is recommended => sedation.
Or 2 / 5 / 10 / 20 / 50 / 100 [L] is selectable.
all
14. Exit.
Jump further with [+] or Exit with [Enter]
all