Instructions

P. 16
Setup and Operation
Version 04-2017
Maintenance:
It is recommended to check once a year if the displayed values are
correct. Two practical check options are:
- Lift the probe above the liquid level. Then check if ~ 0 L is displayed.
- Check the cm value displayed in Step ‘10.Trim height’
(without trimming!).
In case of deviation it is recommended to recalibrate the measuring
probe via menu step 9 or 10. If the problem cannot be fixed the level
probe might be defective. In this case please directly contact Tecson.
New measuring probe (level sensor):
In case of replacement of the level probe it is recommended to call
menu item ‘9.Probe offset’ and execute the item ’Default values!
DANGER
Use in potentially explosive atmospheres not permitted!
May cause explosion or serious injury.
Installation by a specialist company in accordance with the German Industrial Safety
Ordinance!
Installation outside the specified Ex-zone!
ENVIRONMENT & DISPOSAL
To protect the environment, our products must not be disposed of with household
waste.
The product must be disposed of via local collection points or recycling centers.