Instructions
P. 16 
Setup and Operation 
Version 04-2017 
Maintenance: 
It is recommended to check once a year if the displayed values are 
correct. Two practical check options are: 
- Lift the probe above the liquid level. Then check if ~ 0 L is displayed. 
- Check the cm value displayed in Step ‘10.Trim height’  
 (without trimming!). 
In case of deviation it is recommended to recalibrate the measuring 
probe via menu step 9 or 10. If the problem cannot be fixed the level 
probe might be defective. In this case please directly contact Tecson. 
New measuring probe (level sensor): 
In case of replacement of the level probe it is recommended to call 
menu item ‘9.Probe offset’ and execute the item ’Default values’! 
DANGER 
Use in potentially explosive atmospheres not permitted! 
May cause explosion or serious injury. 
Installation by a specialist company in accordance with the German Industrial Safety 
Ordinance! 
Installation outside the specified Ex-zone! 