User's Manual
Removing the transmitter from the Infant
Prior to removing the transmitter from the infant's ankle the transmitter must be placed into the
"Sleep" mode. Failure to place the transmitter into the "Sleep" mode prior to removing the
transmitter from the infant ankle should result in a Cut-band alarm.
To Place the transmitter into the "Sleep" mode.
Press the "On" button on the transmitter tester.
Place the transmitter near the transmitter tester as shown in the picture.(see picture below.)
Enter *23456 (security code) on the keypad on the transmitter tester.
Wait for the tester to display "Enable" within 2 seconds
Press #10001 with the transmitter within 2 inches
Tester will responds with either
"SLEEP OK" or "SLP FAIL" after 10 seconds.
With "Sleep OK" the band should be cut and the Transmitter will discontinue transmission. IF "SLP
FAIL" appears the procedure should be repeated.
Once the transmitter has been placed into the "Sleep" mode, cut the band and remove the cradle
and remaining strap material from the back of the transmitter.
Discard the used strap and cradle.
Clean the transmitter with a mild disinfectant solution. Do not use alcohol or petroleum based
cleaner. Do Not immerse the transmitter in any type of solution with the strap and cradle assembly
removed from the transmitter. (the transmitter is not water or moisture resistant when the strap and
cradle are removed.)
Note: When performing
transmitter testing or placing a
transmitter into Sleep Mode, the
transmitter must be placed near
the testing in the orientation
shown in the picture and the
transmitter must be no more
than 3 inches for testing and no
more than 2 inches for
programming
Cleaning/Disinfecting the Transmitter
The transmitter should be cleaned and disinfected prior to use and between uses. The
transmitter should be cleaned with a mild disinfectant solution.
Do Not Immerse
the
transmitter in solution with the cradle removed. It is recommended that the transmitter
be wiped down with mild disinfectant. If the transmitter requires immersion into
solution then a new cradle should be applied to the transmitter prior to immersion. Do
Not use an autoclave or other type of high temperature device to disinfect the
transmitter.
9 Manual # A20290691 Rev 0