Removing the Transmitter from the Infant
1. Turn on the Tester and verify the green power LED labeled PWR is on solid; if blinking, replace the 9-volt battery.
2. Enter the Security Code *23456 and verify that "ENABLE" is displayed on tester.
3. Orient the transmitter on the front of the infant's leg as shown below.
4. Place the Tester display end so it is actually touching the transmitter front and verify blinking yellow LED labeled
SIGNAL and proper ID code is displayed on the Tester and toggles between ID code and “TX OK”.
5. Enter sleep command on the Tester keypad by pressing #. Hold the tester against the transmitter until the "SLEEP
OK" message is displayed and you hear 5 beeps. This may take up to 15 seconds.
6. You now have 30 seconds to remove the transmitter from the infant by cutting the strap
NOTE:
DO NOT CUT STRAP UNLESS "SLEEP OK "IS BEING DISPLAYED ON THE TESTER.
7. Completely remove the old strap from the transmitter.
NOTE:
FAILURE TO COMPLETELY REMOVE THE OLD STRAP FROM THE
TRANSMITTER MAY RESULT IN FALSE CUTBAND ALARMS.
Cleaning the Transmitter
Secure Care recommends transmitters are cleaned after each use. To properly clean the transmitter use lint free IPA (Isopropyl
Alcohol) wipes. With the transmitter removed from the baby, wipe the transmitter on all sides with the IPA wipe, ensuring
there are no lint fibers attached to the transmitter. When cleaning is finished, store the transmitter in a transmitter containment
box.
NOTE:
ALWAYS STORE TRANSMITTERS IN THE TRANSMITTER CONTAINMENT BOX
WHEN NOT BEING USED ON INFANTS