User's Manual

Connecting the logger and installing it on-site
22
4.1.2 Connecting a hose to the internal pressure sensor
If your logger is equipped with a pressure sensor, a hose can be connected directly to
the connector P
. In order to do so, your hose has to be equipped with a female
quick-fit connector.
NOTE
When connecting a hose to the internal
sensor, the following instructions
must be noted:
Pressure sensors are sensitive to overpressure. Do not exceed the
measuring range in order to avoid irreparable damage to the sensor.
Make sure no dirt or rust can ingress inside the sensors. In case of
doubt, use a filter.
There must not be any water inside the sensor after use. Otherwise,
frost can damage the sensor.
After a sensor has been in use or in storage for a longer period, it must
be cleaned with compressed air.
The sensors are capable of measuring the pressure of both liquid and gaseous media.