User's Manual
Storage / Management of correlations (History)
98
14.2 Managing stored correlations
In the Correlation Management menu the stored correlations are listed. There, the
corresponding measurement data sets can be called up or managed.
To access the menu, proceed as follows:
Step Description
1
In the main menu, tap the button .
2 In the displayed menu, tap the Correlation management button.
Result: The menu opens. The stored correlations are listed
14.2.1 Call up and display a correlation
To call up and display a stored correlation, proceed as follows:
Step Description
3
Select the desired correlation in the list.
4 Tap the Load button.
Result: The measurement data set in question is loaded into the Correlator's
memory (RAM).
5
In the main menu, tap the button .
Result: With the loaded measurement data, a correlation is performed. The
Correlation menu opens and the correlation curve is displayed. The known
analytical tools are available.
Use the Back button or the ESC pushbutton to return to the Start menu.
Element Description
The floppy disk icon indicates that the "History" mode is active. This means
that currently a stored correlation is loaded and ready to be displayed.
In the upper line can be read whether the loaded correlation is an Online
measurement or Offline measurement.
In the lower line the name of the correlation is shown.
If you tap the large button (floppy disk icon), the loaded correlation is displayed again.
Once you use the arrow buttons to set a measuring mode, the History mode is left and
the loaded measurement dataset is removed from the memory.
Call up a correlation