TROUBLESHOOTING
Problem
Probable Causes Try...
No Display
Flow rate always reads “0
Flow rate incorrect
Check for 110 Vac
Test fuse, replace if bad
Check ribbon cable connections
Reverse ow connections
Install ground wire
Install meter in vertical position
Use next smaller ow meter
Check connections on display board
Select another ow meter
Check for proper grounding
Select another ow meter
Install meter in vertical position
No power
Blown fuse
Loose ribbon cable
Reversed ow direction
Missing ground wire
Empty pipe
Flow rate below minimum
Loose wiring or incorrect wiring
Fluid conductivity <20 microSiemens/cm
Missing or incorrect ground wire
Fluid conductivity <20 microSiemens/cm
Empty pipe