Problem
Probable Cause Try...
Laptop screen reads
”No connection” and “Ac-
tive” light is NOT blinking
Laptop screen reads
”No connection,” but “Ac-
tive” light IS blinking
Flow rate incorrect
No ow rate
Wait for Active” light to blink
Replace battery
Tighten cable
Test cable (tech bulletin “DL75/76: Testing the
DC3 Cable” on Downloads page at
www.Seametrics.com)
Make sure K-factor in DL76 matches ow sen-
sor
Mechanical: See if rotor spins freely
EX Magmeter: See if “Status LED” indicator is
blinking
Check for ow rate
DL76 is transmitting data.
Wait two minutes before taking action.
Dead battery in DL76
Loose cable
Bad cable
K-factor setting is incorrect
Damaged ow sensor
Broken wire
TROUBLESHOOTING
CONNECTIONS
Caution: Do not apply external
power to the DL76
+
S
-
DL75
To Flow
Sensor
BATTERY
Sensor Input
Outputs
Sensor Pulse
- +- +
To Pulse Responsive
Metering Pump
FT415
BATTERY
Pulse
Passthru
Pulse
Scaled
Sensor
Input
Power
4-20 mA
To Flow Sensor
S
12-32 Vdc
Loop Power Supply
4-20 mA Device
-
+
+
-
+
-
+
-
+
-
FT420
+
S
-
DL75
BATTERY
+
-
-
DL76 TO FT415
DL76 TO FT420
White
Green
White
Green
Red Black
Black