The Leader in Flow Meter Value CONNECTIONS (In cl ud e sD at es 7/ 19 /0 6 to 11 / 06 ) AR CH IV ED AG2000 Irrigation Magmeter AG2000
INSTALLING THE DC30/DC35 OPTIONAL CABLE Use the diagram below as a guideline for installing the DC30/DC35 optional input/output cable to the AG2000 meter. (NOTE: If you ordered the input/output cable factory-installed, you may disregard this step.) Then proceed to the appropriate wiring diagram for your application on subsequent pages. CAUTION: When the AG2000 is used for AR CH IV ED regulatory purposes, ensure that regulatory requirements are met before breaking the security wire-and-seal.
DL75/AG2000 Pair 1: Serial Output (Technician Use Only) Pair 2: Pulse Output, 30 Vdc max, 10 mA max Pair 3: External Power, 7-24 Vdc, 75 mA min Drain: Connect to earth ground Pair 1-Black AG2000 Cable Blue Pair 1-Red Shielded Burial Cable 3 Twisted Pairs 22 AWG Stranded Battery +S - Pair 2-Black Pair 2-Red Orange AG2000 Pair 3-Black Pair 3-Red Not Used White *Colored Sleeves readily identify the 3 pairs. If the cable needs to be cut back, the pair identification is printed on each wire.
FT420/DL75/AG2000 WITH OR WITHOUT EXTERNAL POWER 2 Wire NonShielded Cable Pair 1: Serial Output (Technician Use Only) Pair 2: Pulse Output, 30 Vdc max, 10 mA max Pair 3: External Power, 7-24 Vdc, 75 mA min Drain: Connect to earth ground PULSE PASSTHRU Power 4-20 mA Not Used Pair 1-Black AG2000 Cable Blue Pair 1-Red Shielded Burial Cable 3 Twisted Pairs 22 AWG Stranded Pair 2-Red Orange AG2000 AR CH IV ED PULSE SCALED Sensor Input Pair 2-Black Pair 3-Black Pair 3-Red Not Used White *Colore