User`s manual
Table Of Contents
- SEASAVE
- Limited Liability Statement
- Table of Contents
- Section 1: Introduction
- Section 2: Installation & Use
- Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
- Introduction
- Instrument Configuration
- Viewing, Modifying, or Creating .con File
- SBE 9plus Configuration
- SBE 16 SEACAT C-T Recorder Configuration
- SBE 16plus SEACAT C-T Recorder Configuration
- SBE 19 SEACAT Profiler Configuration
- SBE 19plus SEACAT Profiler Configuration
- SBE 21 Thermosalinograph Configuration
- SBE 25 SEALOGGER Configuration
- SBE 45 MicroTSG Configuration
- SBE 49 FastCAT Configuration
- Section 4: Configure Inputs, Part II - Calibration Coefficients
- Accessing Calibration Coefficients Dialog Boxes
- Calibration Coefficients for Frequency Sensors
- Calibration Coefficients for A/D Count Sensors
- Calibration Coefficients for Voltage Sensors
- Pressure (Strain Gauge) Calibration Coefficients
- Altimeter Calibration Coefficients
- Fluorometer Calibration Coefficients
- Methane Sensor Calibration Coefficients
- OBS/Nephelometer Calibration Coefficients
- Oxidation Reduction Potential (ORP) Calibration Coefficients
- Oxygen Calibration Coefficients
- PAR/Irradiance Calibration Coefficients
- pH Calibration Coefficients
- Pressure/FGP (voltage output) Calibration Coefficients
- Suspended Sediment Calibration Coefficients
- Transmissometer Calibration Coefficients
- User Polynomial (for user-defined sensor) Calibration Coefficients
- Zaps Calibration Coefficients
- Section 5: Configure Inputs, Part III – Serial Ports, Water Sampler, TCP/IP Ports, Miscellaneous, & Pump Control
- Section 6: Configure Outputs
- Section 7: Display - Setting Up SEASAVE Displays
- Section 8: Real-Time Data & Real-Time Control - Real-Time Data Acquisition
- Section 9: Archived Data Displaying Archived Data
- Section 10: Processing Data
- Appendix I: Command Line Operation
- Appendix II: Configure (.con) File Format
- Appendix III: Software Problems
- Appendix IV: Derived Parameter Formulas
- Index

Section 8: Real-Time Data and Real-Time Control - Real-Time Data Acquisition
85
Firing Bottles
Water sampler bottles can be fired by command from SEASAVE. SEASAVE
automatically writes bottle sequence number, bottle position, date, time, and
beginning and ending scan numbers to a bottle log (.bl) file each time a bottle
fire confirmation is received from the water sampler. The beginning and
ending scan numbers correspond to approximately a 1.5-second duration for
each bottle. For a 911plus system, SEASAVE also automatically sets the
bottle confirm bit in the data (.hex) file for all scans within a 1.5-second
duration after a bottle firing confirmation is received from the water sampler.
To fire bottles:
1. Set up the water sampler in Configure Inputs (see Water Sampler in
Section 5: Configure Inputs, Part III – Serial Ports, Water Sampler,
TCP/IP Ports, Miscellaneous, and Pump Control).
2. In the Real-Time Control menu, select Fire Bottle Control. The Bottle Fire
dialog box appears (you can leave this open throughout the cast).
3. Start real-time data acquisition.
4. If you selected Sequential or Table driven in the Water Sampler setup
(Step 1), the Bottle Fire dialog box displays the number of the next bottle
to be fired. If you selected User Input in the Water Sampler setup (Step 1),
select the bottle you want to fire next.
When desired, click Fire Bottle.
When SEASAVE receives a bottle fired confirmation from a Carousel, it:
• (for all CTDs) Writes a line to an output file (same filename as the data
file) with a .bl extension. The .bl file contains the bottle firing sequence
number, bottle position, date, time, and beginning and ending scan
number (to provide 1.5 seconds of scans) for the fired bottle.
• (for SBE 911plus only) Sets a bit in the modulo word high for
1.5 seconds and increments the number of bottles fired. Later, when the
raw data file is converted in SBE Data Processing’s Data Conversion
module, scans with this bottle confirmation bit are written to a file with
a .ros extension.
Note:
The .bl file has the same path and
file name as the data file. For
example, if the data file is
c:\test1.hex, the .bl file is c:\test1.bl.
Notes:
• The Fire Bottle button in the Bottle
Fire dialog box is inaccessible until
you start saving data to a file. So, if
you did not select Store data on disk
(or selected Delay storing data to
disk until Start Archiving command
is sent and did not yet send the Start
Archiving command), the Fire Bottle
button is grayed out.
• If desired, you can fire bottles
without using the Bottle Fire dialog
box. Each time you want to fire a
bottle, press Ctrl F3.
• If Auto Fire firing sequence was
selected on the Water Sampler tab
in Configure Inputs, the Fire Bottle
Control dialog box is inaccessible.