Section 6: Configure Outputs
58
Section 6: Configure Outputs
This section describes the setup of the following in Configure Outputs:
Serial data output
Serial ports
Shared file output
Mark variables
TCP/IP output
TCP/IP ports
SBE 11plus alarms (only applicable if instrument configuration [.con] file
is for 911plus/917plus CTD)
SBE 14 Remote Display
Header form
Diagnostics
Some outputs – serial data out, shared file out, SBE 11plus alarms, and
remote display - can be reconfigured during data acquisition without
interrupting data acquisition. For example, if you start a cast and realize that
you forgot to set up the serial data output, you can select Configure Outputs,
and make and save the desired changes, all without interrupting the data
acquisition. Once the desired changes are saved, the serial data will begin to
output to the desired COM port (of course, any data that was acquired before
you modified the setup will not be output in the serial data stream).
For setup of the inputs, see Section 3: Configure Inputs, Part I - Instrument
.con File, Section 4: Configure Inputs, Part II – Calibration Coefficients,
and Section 5: Configure Inputs, Part III – Serial Ports, Water Sampler,
TCP/IP Ports, Miscellaneous, and Pump Control.
Note:
Setup of all parameters (except
Diagnostics) in Configure Outputs
is included in the SEASAVE
program setup (.psa) file. To save
the setup, you must save the .psa
file (File menu / Save Setup File)
before exiting SEASAVE.