User`s manual
Table Of Contents
- SEASAVE
- Limited Liability Statement
- Table of Contents
- Section 1: Introduction
- Section 2: Installation & Use
- Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
- Introduction
- Instrument Configuration
- Viewing, Modifying, or Creating .con File
- SBE 9plus Configuration
- SBE 16 SEACAT C-T Recorder Configuration
- SBE 16plus SEACAT C-T Recorder Configuration
- SBE 19 SEACAT Profiler Configuration
- SBE 19plus SEACAT Profiler Configuration
- SBE 21 Thermosalinograph Configuration
- SBE 25 SEALOGGER Configuration
- SBE 45 MicroTSG Configuration
- SBE 49 FastCAT Configuration
- Section 4: Configure Inputs, Part II - Calibration Coefficients
- Accessing Calibration Coefficients Dialog Boxes
- Calibration Coefficients for Frequency Sensors
- Calibration Coefficients for A/D Count Sensors
- Calibration Coefficients for Voltage Sensors
- Pressure (Strain Gauge) Calibration Coefficients
- Altimeter Calibration Coefficients
- Fluorometer Calibration Coefficients
- Methane Sensor Calibration Coefficients
- OBS/Nephelometer Calibration Coefficients
- Oxidation Reduction Potential (ORP) Calibration Coefficients
- Oxygen Calibration Coefficients
- PAR/Irradiance Calibration Coefficients
- pH Calibration Coefficients
- Pressure/FGP (voltage output) Calibration Coefficients
- Suspended Sediment Calibration Coefficients
- Transmissometer Calibration Coefficients
- User Polynomial (for user-defined sensor) Calibration Coefficients
- Zaps Calibration Coefficients
- Section 5: Configure Inputs, Part III – Serial Ports, Water Sampler, TCP/IP Ports, Miscellaneous, & Pump Control
- Section 6: Configure Outputs
- Section 7: Display - Setting Up SEASAVE Displays
- Section 8: Real-Time Data & Real-Time Control - Real-Time Data Acquisition
- Section 9: Archived Data Displaying Archived Data
- Section 10: Processing Data
- Appendix I: Command Line Operation
- Appendix II: Configure (.con) File Format
- Appendix III: Software Problems
- Appendix IV: Derived Parameter Formulas
- Index

Section 4: Configure Inputs, Part II - Calibration Coefficients
46
PAR/Irradiance Calibration Coefficients
Underwater PAR Sensor
Enter M, B, calibration constant, multiplier, and offset.
PAR = [multiplier * (10
9
* 10
(V-B) / M
) / calibration constant] + offset
where
calibration constant, M, and B are dependent on sensor type;
multiplier = 1.0 for output units of μEinsteins/m
2
sec
• Biospherical PAR sensor
- PAR sensor with built-in log amplifier (QSP-200L, QCP-200L,
QSP-2300L, QCP-2300L, or MCP-2300)]:
Typically, M = 1.0 and B = 0.0.
Calibration constant
= 10
5
/ wet calibration factor from Biospherical calibration sheet.
- PAR sensor without built-in log amplifier (QSP-200PD, QSP-2200 (PD),
or QCP 2200 (PD)):
M and B are taken from Sea-Bird calibration sheet.
Calibration constant
= C
S
calibration coefficient from Sea-Bird calibration sheet
= 6.022 x 10
13
/ C
w
from Biospherical calibration sheet
• LI-COR PAR sensor
Calibration constant is in water calibration constant (in units of
μamps/1000 μmoles/m
2.
sec) from Licor or Sea-Bird calibration sheet.
M and B are taken from Sea-Bird calibration sheet.
• Chelsea PAR sensor
Calibration constant = 10
9
/ 0.046
M = 1.0 / (log e * A1 * 1000) = 1.0 / (0.43429448 * A1 * 1000)
B = - M * log e * A0 = - A0 / (A1 * 1000)
where A0 and A1 are constants from Chelsea calibration sheet with an
equation of form: PAR = A0 + (A1 * mV)
Note: SEASAVE can process data for an instrument interfacing with up to two
underwater PAR/irradiance sensors.
Biospherical Surface PAR Sensor
A surface PAR sensor is selected by clicking Surface PAR voltage added in
the Configure dialog box. Enter conversion factor and ratio multiplier.
Notes:
• See Application Note 11General
for multiplier values for output units
other than μEinsteins/m
2
sec.
• See Application Notes 11QSP-L
(Biospherical sensor with built-in
log amplifier), 11QSP-PD
(Biospherical sensor without built-
in log amplifier), 11Licor (LI-COR
sensor), and 11Chelsea for
complete description of calculation
of calibration coefficients for
underwater PAR sensors.
•
Selection of Par / Irradiance,
Biospherical / Licor as the voltage
sensor is also applicable to the
Chelsea PAR sensor.
Notes:
• See Application Note 11 General
for conversion factor values for
output units other than
μEinsteins/m
2
sec.
• For complete description of
calculation of calibration
coefficients for surface PAR,
see Application Note 11S
(SBE 11plus Deck Unit) or
47 (SBE 33 or 36 Deck Unit).