User`s manual
Table Of Contents
- SEASAVE
- Limited Liability Statement
- Table of Contents
- Section 1: Introduction
- Section 2: Installation & Use
- Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
- Introduction
- Instrument Configuration
- Viewing, Modifying, or Creating .con File
- SBE 9plus Configuration
- SBE 16 SEACAT C-T Recorder Configuration
- SBE 16plus SEACAT C-T Recorder Configuration
- SBE 19 SEACAT Profiler Configuration
- SBE 19plus SEACAT Profiler Configuration
- SBE 21 Thermosalinograph Configuration
- SBE 25 SEALOGGER Configuration
- SBE 45 MicroTSG Configuration
- SBE 49 FastCAT Configuration
- Section 4: Configure Inputs, Part II - Calibration Coefficients
- Accessing Calibration Coefficients Dialog Boxes
- Calibration Coefficients for Frequency Sensors
- Calibration Coefficients for A/D Count Sensors
- Calibration Coefficients for Voltage Sensors
- Pressure (Strain Gauge) Calibration Coefficients
- Altimeter Calibration Coefficients
- Fluorometer Calibration Coefficients
- Methane Sensor Calibration Coefficients
- OBS/Nephelometer Calibration Coefficients
- Oxidation Reduction Potential (ORP) Calibration Coefficients
- Oxygen Calibration Coefficients
- PAR/Irradiance Calibration Coefficients
- pH Calibration Coefficients
- Pressure/FGP (voltage output) Calibration Coefficients
- Suspended Sediment Calibration Coefficients
- Transmissometer Calibration Coefficients
- User Polynomial (for user-defined sensor) Calibration Coefficients
- Zaps Calibration Coefficients
- Section 5: Configure Inputs, Part III – Serial Ports, Water Sampler, TCP/IP Ports, Miscellaneous, & Pump Control
- Section 6: Configure Outputs
- Section 7: Display - Setting Up SEASAVE Displays
- Section 8: Real-Time Data & Real-Time Control - Real-Time Data Acquisition
- Section 9: Archived Data Displaying Archived Data
- Section 10: Processing Data
- Appendix I: Command Line Operation
- Appendix II: Configure (.con) File Format
- Appendix III: Software Problems
- Appendix IV: Derived Parameter Formulas
- Index

Section 4: Configure Inputs, Part II - Calibration Coefficients
37
Pressure (Paroscientific Digiquartz) Calibration Coefficients
Enter the sets of C, D, and T coefficients from the calibration sheet. Enter zero
for any higher-order coefficients that are not listed on the calibration sheet.
Enter values for slope (default = 1.0; do not change unless sensor has been
recalibrated) and offset (default = 0.0) to make small corrections for
sensor drift.
• For the SBE 9plus, also enter AD590M and AD590B coefficients from
the configuration sheet.
Bottles Closed (HB - IOW) Calibration Coefficients
No calibration coefficients are entered for this parameter.
The number of bottles closed is calculated by SBE Data Processing’s Data
Conversion module based on frequency range.
Sound Velocity (IOW) Calibration Coefficients
Enter coefficients a0, a1, and a2.
Value = a0 + a1 * frequency + a2 * frequency
2
Note:
See Calibration Coefficients for
A/D Count Sensors below for
information on strain gauge
pressure sensors used on the
SBE 16plus, 19plus, and 49.
See Calibration Coefficients for
Voltage Sensors below for
information on strain gauge
pressure sensors used on
other instruments.