User`s manual

Table Of Contents
Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
29
Shown below is an example status (DS) response in SEATERM that
corresponds to the setup shown in the Configuration dialog box above. Shown
below the appropriate lines are the commands used in SEATERM to modify
the setup of parameters critical to use of the instrument with SEASAVE, as
well as any explanatory information.
SEACAT THERMOSALINOGRAPH V4.2a SERIAL NO. 4300 05/15/2003
14:23:14
ioper = 50.7 ma, vmain = 11.4, vlith = 8.8,
iext01 = 76.2 ma
samples = 0, free = 1396736
sample interval = 5 seconds
(Sample interval [SI] must match Sample interval seconds in .con file.)
sample external SBE 38 temperature sensor
(External temperature sensor [SBE38=, SBE3=] must match Remote temperature
in .con file.)
no. of volts sampled = 1
(Number of auxiliary voltage sensors enabled [SVx] must match External voltage
channels in .con file.)
output format = SBE21
(Output format must be set to SBE 21 [F1] to acquire data in SEASAVE.)
start sampling when power on = yes
average data during sample interval = yes
logging data = no
voltage cutoff = 7.5 volts