Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
25
Shown below is an example status (DS) response in SEATERM that
corresponds to the setup shown in the Configuration dialog box above. Shown
below the appropriate lines are the commands used in SEATERM to modify
the setup of parameters critical to use of the instrument with SEASAVE, as
well as any explanatory information.
SEACAT PROFILER V3.1B SN 936 02/10/94 13:33:23.989
strain gauge pressure sensor: S/N = 12345,
range = 1000 psia, tc = 240
(Pressure sensor (strain gauge or Digiquartz) must match Pressure sensor type in
.con file.)
clk = 32767.766 iop = 172 vmain = 8.1 vlith = 5.8
mode = PROFILE ncasts = 0
(Mode must be profile [MP] if setting up .con file for SBE 19; create .con file for
SBE 16 for SBE 19 in moored mode [MM].)
sample rate = 1 scan every 0.5 seconds
(Sample rate [SR] must match 0.5 second intervals in .con file.)
minimum raw conductivity frequency for pump turn on =
3206 hertz
pump delay = 40 seconds
samples = 0 free = 174126 lwait = 0 msec
battery cutoff = 7.2 volts
number of voltages sampled = 2
(Number of auxiliary voltage sensors enabled [SVn] must match External voltage
channels in .con file.)
logdata = NO