User`s manual

Table Of Contents
Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
23
Shown below is an example status (DS) response in SEATERM that
corresponds to the setup shown in the Configuration dialog box above. Shown
below the appropriate lines are the commands used in SEATERM to modify
the setup of parameters critical to use of the instrument with SEASAVE, as
well as any explanatory information.
SBE 16plus V 1.6e SERIAL NO. 4300 03 Mar 2005 14:11:48
vbatt = 10.3, vlith = 8.5, ioper = 62.5 ma,
ipump = 21.6 ma, iext01 = 76.2 ma, iserial = 48.2 ma
status = not logging
sample interval = 10 seconds, number of measurements
per sample = 2
(Sample interval [SAMPLEINTERVAL=] must match Sample interval seconds in
.con file.)
samples = 823, free = 465210
run pump during sample, delay before sampling =
2.0 seconds
transmit real-time = yes
(Real-time data transmission must be enabled [TXREALTIME=Y] to acquire
data in SEASAVE.)
battery cutoff = 7.5 volts
pressure sensor = strain gauge, range = 1000.0
(Internal pressure sensor [PTYPE=] must match Pressure sensor type in .con file.)
SBE 38 = yes, SBE 50 = no, Gas Tension Device = no
(Selection/enabling of RS-232 sensors [SBE38=, SBE50=, GTD=, DUALGTD=]
must match Serial RS-232C sensor in .con file.)
Ext Volt 0 = yes, Ext Volt 1 = yes, Ext Volt 2 = no, Ext
Volt 3 = no
(Number of external voltage sensors enabled [VOLT0= through VOLT3=] must
match External voltage channels in .con file.)
echo commands = yes
output format = raw HEX
(Output format must be set to raw Hex [OUTPUTFORMAT=0] to acquire data in
SEASAVE.)
serial sync mode disabled
(Serial sync mode must be disabled [SYNCMODE=N] to acquire data in
SEASAVE.)