Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
20
Shown below is an example status (DS) response in SEATERM that
corresponds to the setup shown in the Configuration dialog box above. Shown
below the appropriate lines are the commands used in SEATERM to modify
the setup of parameters critical to use of the instrument with SEASAVE, as
well as any explanatory information.
SBE 11plus V 5.1f
Number of scans to average = 1
(11plus reads this from .con file in SEASAVE after Deck Unit is reset.)
pressure baud rate = 9600
NMEA baud rate = 4800
surface PAR voltage added to scan
(Enabling of surface PAR [ADDSPAR=] must match Surface PAR voltage added in
.con file.)
A/D offset = 0
GPIB address = 1
(GPIB address must be 1 [GPIB=1] to use SEASAVE, if Computer interface is
IEEE-488 (GPIB) in .con file.)
advance primary conductivity 0.073 seconds
advance secondary conductivity 0.073 seconds
autorun on power up is disabled