User`s manual
Table Of Contents
- SEASAVE
- Limited Liability Statement
- Table of Contents
- Section 1: Introduction
- Section 2: Installation & Use
- Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
- Introduction
- Instrument Configuration
- Viewing, Modifying, or Creating .con File
- SBE 9plus Configuration
- SBE 16 SEACAT C-T Recorder Configuration
- SBE 16plus SEACAT C-T Recorder Configuration
- SBE 19 SEACAT Profiler Configuration
- SBE 19plus SEACAT Profiler Configuration
- SBE 21 Thermosalinograph Configuration
- SBE 25 SEALOGGER Configuration
- SBE 45 MicroTSG Configuration
- SBE 49 FastCAT Configuration
- Section 4: Configure Inputs, Part II - Calibration Coefficients
- Accessing Calibration Coefficients Dialog Boxes
- Calibration Coefficients for Frequency Sensors
- Calibration Coefficients for A/D Count Sensors
- Calibration Coefficients for Voltage Sensors
- Pressure (Strain Gauge) Calibration Coefficients
- Altimeter Calibration Coefficients
- Fluorometer Calibration Coefficients
- Methane Sensor Calibration Coefficients
- OBS/Nephelometer Calibration Coefficients
- Oxidation Reduction Potential (ORP) Calibration Coefficients
- Oxygen Calibration Coefficients
- PAR/Irradiance Calibration Coefficients
- pH Calibration Coefficients
- Pressure/FGP (voltage output) Calibration Coefficients
- Suspended Sediment Calibration Coefficients
- Transmissometer Calibration Coefficients
- User Polynomial (for user-defined sensor) Calibration Coefficients
- Zaps Calibration Coefficients
- Section 5: Configure Inputs, Part III – Serial Ports, Water Sampler, TCP/IP Ports, Miscellaneous, & Pump Control
- Section 6: Configure Outputs
- Section 7: Display - Setting Up SEASAVE Displays
- Section 8: Real-Time Data & Real-Time Control - Real-Time Data Acquisition
- Section 9: Archived Data Displaying Archived Data
- Section 10: Processing Data
- Appendix I: Command Line Operation
- Appendix II: Configure (.con) File Format
- Appendix III: Software Problems
- Appendix IV: Derived Parameter Formulas
- Index

Section 3: Configure Inputs, Part I - Instrument Configuration (.con file)
18
All Instrument Configuration dialog boxes include:
• List of instrument configuration options at the top (instrument-
specific), such as number of auxiliary channels, pressure sensor type,
and addition of Surface PAR and NMEA to the CTD data string.
• Channel/Sensor Table: This table reflects the options selected at the
top (for example, the number of voltage sensors listed in the table
agrees with the user-selection for External voltage channels). Shaded
sensors cannot be removed or changed to another type of sensor. All
others are optional.
¾ To change a sensor type and input its calibration coefficients:
After you specify the number of frequency and/or voltage
channels at the top of the dialog box, click a (non-shaded) sensor
and click Select to pick a different sensor for that channel
(or right click on the sensor and select Select New . . Sensor).
A dialog box with a list of sensors appears.
Double click on the desired sensor. The Calibration Coefficients
dialog box appears. An example is shown below for a pH sensor:
Enter the desired values and click OK.
¾ To change a sensor’s calibration coefficients:
In the Configuration dialog box, click a sensor and click Modify
to change the calibration coefficients for that sensor (or right
click on the sensor and select Modify . . Calibration, or double
click on the sensor). The Calibration Coefficients dialog box
appears (example shown above). See Section 4: Configure
Inputs, Part II - Calibration Coefficients for calculation
of coefficients.