Specifications

4
Non-Functioning or Poorly Functioning SBE 5T
Perform the inspection procedures listed above in Yearly Maintenance for SBE 5T. If you do not discover the problem
there, proceed as follows.
1. Unscrew the connector end cap retaining ring. Pull out the
end cap and attached electronics from the housing.
2. Verify that the magnet can spin freely and is not broken or
damaged.
3. Look for other signs of damage on the electronics.
4. Inspect the connector end cap o-ring and the mating surface in the housing for dirt, nicks, and cuts. Clean as
necessary. If the o-ring or mating surface is damaged, return the pump to Sea-Bird for repairs.
Sea-Bird recommends that connector end cap o-ring replacement be performed at the factory, because the
pumps physical configuration makes customer-replacement of this o-ring difficult to perform without
special tools.
5. Apply a light coat of o-ring lubricant (Parker Super O Lube) to the o-ring and mating surfaces. Gently place a new
desiccant bag on the electronics (see Application Note 71 for desiccant use and regeneration). Reinstall the
electronics in the housing, until the o-ring has fully seated. Reinstall the retaining ring on the connector end cap.
Retaining ring
Connector
end cap
Retaining ring
Desiccant
Magnet
Housing
PN 30082
Connector end
cap o-ring
