User`s manual
Manual revision 016 Section 5: Routine Maintenance and Calibration SBE 49
52
Anti-Foulant Device in Pump Exhaust Tubing
1. Carefully cut the cable tie securing the Tygon tubing to the anti-foulant
device cap. Slip the Tygon tubing off of the anti-foulant device cap.
2. Unscrew the cap with a socket wrench.
3. Remove the old Anti-Foulant Device. If the old device is difficult to
remove, use needle-nose pliers and carefully break up material.
4. Place the new Anti-Foulant Device in the cup.
5. Rethread the cap onto the cup. Do not over tighten.
6. Slip the Tygon tubing back onto the cap. Secure with a new cable tie.
Cut cable tie and slip
Tygon tubing off of anti
-
foulant device cap
Unscrew anti-foulant
device cap from anti
-
foulant device cup
Anti-foulant device
cup holds Anti
-
Foulant Device