Manual revision 016 Section 4: Deploying and Operating FastCAT SBE 49
35
OutputFormat=2 (raw data in decimal)
Data is output in the order listed, with a comma followed by a space between
each parameter. Shown with each parameter are the number of digits and the
placement of the decimal point. Leading zeros are suppressed, except for one
zero to the left of the decimal point.
1. Temperature A/D counts = tttttt
2. Conductivity frequency (Hz) = cccc.ccc
3. Pressure sensor pressure A/D counts = pppppp
4. Pressure sensor temperature compensation voltage = v.vvvv
OutputFormat=3 (engineering units in decimal)
Data is output in the order listed, with a comma followed by a space between
each parameter. Shown with each parameter are the number of digits and the
placement of the decimal point. Leading zeros are suppressed, except for one
zero to the left of the decimal point.
Note that if ProcessRealTime=Y, the output data has been aligned, filtered,
and corrected for cell thermal mass effects.
1. Temperature (°C, ITS-90) = ttt.tttt
2. Conductivity (S/m) = cc.ccccc
3. Pressure (decibars) = pppp.ppp
4. Salinity (psu) = sss.ssss (if OutputSal=Y)
5. Sound velocity (meters/second) = vvvv.vvv (if OutputSV=Y)
Example: example scan = tttttt, cccc.ccc, pppppp, v.vvvv = 676721, 7111.133, 791745, 2.4514
Temperature = tttttt = 676721;
temperature A/D counts = 676721
Conductivity = cccc.ccc = 7111.133;
conductivity frequency = 7111.133 Hz
Pressure = pppppp = 791745;
Pressure A/D counts = 791745
Pressure temperature compensation = v.vvvv = 2.4514;
Pressure temperature = 2.4514 volts
Example: FastCAT with OutputSal=N, OutputSV=N
example scan = ttt.tttt, cc.ccccc, pppp.ppp = 23.7658, 0.00019, 0.062
Temperature = ttt.tttt = 23.7658;
temperature (°C, ITS-90) = 23.7658
Conductivity = cc.ccccc = 0.00019;
conductivity (S/m) = 0.00019
Pressure = pppp.ppp = 0.062;
pressure (decibars) = 0.062