User`s manual

Manual revision 016 Section 4: Deploying and Operating FastCAT SBE 49
33
Coefficients Commands
DCal Display calibration coefficients.
Equivalent to Coefficients on Toolbar.
Use the commands listed below to modify a particular coefficient or date:
Temperature
TCalDate=S S=calibration date
TA0=F F=A0
TA1=F F=A1
TA2=F F=A2
TA3=F F=A3
TOffset=F F=offset correction
Conductivity
CCalDate=S S=calibration date
CG=F F=G
CH=F F=H
CI=F F=I
CJ=F F=J
CPCOR=F F=pcor
CTCOR=F F=tcor
CSlope=F F=slope correction
Pressure
PCalDate=S S=calibration date
PRange=F F=sensor full scale range (psia)
POffset=F F=offset correction (decibars)
PA0=F F=A0
PA1=F F=A1
PA2=F F=A2
PTempA0=F F=pressure temperature A0
PTempA1=F F=pressure temperature A1
PTempA2=F F=pressure temperature A2
PTCA0=F F=pressure temperature compensation ptca0
PTCA1=F F=pressure temperature compensation ptca1
PTCA2=F F=pressure temperature compensation ptca2
PTCB0=F F=pressure temperature compensation ptcb0
PTCB1=F F=pressure temperature compensation ptcb1
PTCB2=F F=pressure temperature compensation ptcb2
Example: Display coefficients for FastCAT (user input in bold).
S>dcal
SBE 49 FastCAT V 1.3a SERIAL NO. 0055
temperature: 26-apr-01
TA0 = -3.178124e-06
TA1 = 2.751603e-04
TA2 = -2.215606e-06
TA3 = 1.549719e-07
TOFFSET = 0.000000e+00
conductivity: 26-apr-01
G = -9.855242e-01
H = 1.458421e-01
I = -3.290801e-04
J = 4.784952e-05
CPCOR = -9.570000e-08
CTCOR = 3.250000e-06
CSLOPE = 1.000000e+00
Pressure S/N = 023674, range = 1000 psia: 25-apr-01
PA0 = -6.893561e-01
PA1 = 1.567975e-02
PA2 = -6.637727e-10
PTCA0 = 5.246558e+05
PTCA1 = -4.886082e+00
PTCA2 = 1.257885e-01
PTCB0 = 2.489275e+01
PTCB1 = -8.500000e-04
PTCB2 = 0.000000e+00
PTEMPA0 = -6.634546e+01
PTEMPA1 = 5.093069e+01
PTEMPA2 = 1.886260e-01
POFFSET = 0.000000e+00
Note:
F = floating point number
S = string with no spaces