User`s manual

Manual revision 016 Section 4: Deploying and Operating FastCAT SBE 49
32
Polled Sampling Commands
PumpOn Turn pump on. Use this command:
Before sending TS to obtain pumped
conductivity data, or
To test pump.
PumpOff Turn pump off.
TS Take 1 sample and transmit data.
Testing Commands
The FastCAT takes and transmits data for 100 samples for each test. Press the
Esc key or Stop on the Toolbar to stop a test.
TT Measure temperature, transmit
converted data.
TC Measure conductivity, transmit
converted data.
TP Measure pressure, transmit converted data.
TTR Measure temperature, transmit raw data.
TCR Measure conductivity, transmit raw data.
TPR Measure pressure, transmit raw data.
Note:
Testing commands do not
automatically turn the pump on. Thus,
they report conductivity from
essentially the same sample of water
for all 100 samples, because the
pump does not run and water does
not freely flow through the conductivity
cell. To get conductivity from fresh
samples, send PumpOn before
sending TC or TCR, and then send
PumpOff
when the test is complete.