User`s manual

Manual revision 016 Section 4: Deploying and Operating FastCAT SBE 49
31
Autonomous Sampling Commands (continued)
ProcessRealTime=x x= Y: Apply alignment, filtering, and
conductivity cell thermal mass corrections
to real-time data. Only applies if
OutputFormat=1 or 3. Default.
x= N: Do not apply corrections to data.
TAdvance=x x= Time to advance temperature data
relative to conductivity and pressure data.
Range 0 to 0.125 seconds; default
0.0625 seconds. Only applies if
ProcessRealTime=Y and
OutputFormat=1 or 3.
Alpha=x x= Conductivity cell thermal mass alpha
correction. Range 0.02 to 0.05; default
0.03. Only applies if ProcessRealTime=Y
and OutputFormat=1 or 3.
Tau=x x= Conductivity cell thermal mass tau
correction. Range 5.0 to 10.0; default 7.0.
Only applies if ProcessRealTime=Y and
OutputFormat=1 or 3.
AutoRun=x x=N: Wait for a command when power is
applied. Default.
x=Y: Start autonomous sampling
automatically when power is applied.
Start Start autonomous sampling now
(applicable if AutoRun=N or if you just
sent AutoRun=Y).
Stop Stop autonomous sampling.
Press Enter key to get S> prompt
before entering Stop.
Notes:
You may need to send Stop several
times to get the FastCAT to respond.
You can also stop autonomous
sampling by clicking the Stop key on
the Toolbar, typing Ctrl Z, or
removing power.
Note:
After you send AutoRun=Y,
to start sampling immediately:
turn power off and then on again, or
send Start.
Notes:
See Real-Time Data Corrections for
detailed explanations of
ProcessRealTime=, TAdvance=,
Alpha=, and Tau=.
If ProcessRealTime=Y but
OutputFormat=0 or 2 (output raw
hex or raw decimal data), the status
reply shows:
real-time temperature and
conductivity correction
enabled, not applied to raw
data,
and the FastCAT does not perform
real-time corrections.