User`s manual
Manual revision 016 Section 4: Deploying and Operating FastCAT SBE 49
28
Status Command
DS Display operating status and
setup parameters.
Equivalent to Status on Toolbar.
List below includes, where applicable,
command used to modify parameter.
• firmware version, serial number
• number of scans to average [NAvg=]
• pressure sensor type and
range [PRange=]
• minimum conductivity frequency for
pump turn-on [MinCondFreq=] and
pump turn-on delay [PumpDelay=]
• start autonomous sampling
automatically when power applied
[AutoRun=]?
• output format [OutputFormat=]
• output salinity [OutputSal=] and
sound velocity [OutputSV=] with
each sample? (only if
OutputFormat=3)
• time to advance temperature to align
with conductivity [TAdvance=]
• conductivity cell thermal mass alpha
correction [Alpha=]
• conductivity cell thermal mass tau
correction [Tau=]
• enable real-time corrections
[ProcessRealTime=]?
Example: Status (DS) command (user input in bold; command used to modify parameter in parentheses)
S>DS
SBE 49 FastCAT V 1.3a SERIAL NO. 0055
number of scans to average = 1 [NAvg=]
pressure sensor = strain gauge, range = 1000.0 [PRange=]
minimum cond freq = 3000, pump delay = 30 sec [MinCondFreq=, PumpDelay=]
start sampling on power up = no [AutoRun=]
output format = converted decimal [OutputFormat=]
output salinity = yes, output sound velocity = yes [OutputSal=, OutputSV=]
temperature advance = 0.0625 seconds [TAdvance=]
celltm alpha = 0.03 [Alpha=]
celltm tau = 7.0 [Tau=]
real-time temperature and conductivity correction enabled for converted data [ProcessRealTime=]
Note:
If ProcessRealTime=Y but
OutputFormat=0 or 2 (output raw
hexadecimal or raw decimal data), the
status reply shows:
real-time temperature and
conductivity correction
enabled, not applied to raw
data,
and the FastCAT does not perform
real-time corrections.