User`s manual

Manual revision 016 Section 4: Deploying and Operating FastCAT SBE 49
21
Polled Sampling
On command (TS), the FastCAT takes one sample and transmits the
data real-time.
The pump does not turn on automatically for polled sampling
(MinCondFreq= and PumpDelay= do not apply). To run the pump before
taking a sample, send PumpOn to turn the pump on before sending TS. Send
PumpOff to turn the pump off after taking the sample.
Example: Polled Sampling (user input in bold)
Apply power and wake up FastCAT. Set up to not start automatic
autonomous sampling when power is applied, to output data in
decimal engineering units format, and to output salinity and sound
velocity as well as conductivity, temperature, and pressure.
Verify setup with status command. Remove power.
(Apply power, then click Connect on Toolbar to wake up.)
S>AUTORUN=N
S>OUTPUTFORMAT=3
S>OUTPUTSAL=Y
S>OUTPUTSV=Y
S>DS (to verify setup)
(Remove power.)
When ready to take a sample, put FastCAT in water. Apply power
and wake up FastCAT. Command FastCAT to turn pump on, take
a sample, and turn pump off. Remove power. Repeat as desired.
(Apply power, then click Connect on Toolbar to wake up.)
S>PUMPON
S>TS
S>PUMPOFF
(Remove power.)