User`s manual

Section 7: Real-Time Data Acquisition
70
Turning Pump On / Off
SEASAVE allows you to manually turn a 911plus pump on and off during
data acquisition. This may be useful if your system is integrated with an
acoustic instrument, to provide a quiet period during its data acquisition.
To manually turn SBE 911plus pump on / off:
1. Start SEASAVE from the command line (select Run in the Windows Start
menu), using the -pc command line option:
Path\seasave.exe -pc (path is location of seasave.exe on your computer)
2. SEASAVE opens. Set up the system and displays as desired.
3. Start real-time data acquisition.
4. When desired:
In the RealTime Data menu, select Turn Pump On or Turn Pump Off, OR
Press Ctrl F2 (pump on) or Ctrl F4 (pump off).
Note:
You must start SEASAVE from
the command line, using the -pc
command, to enable pump turn
on / off from SEASAVE. If you do
not, Turn Pump On and Turn
Pump Off will remain grayed out
and be unavailable after you start
data acquisition.