User`s manual

Section 7: Real-Time Data Acquisition
68
Firing Bottles
Water sampler bottles can be fired by command from SEASAVE. SEASAVE
automatically writes bottle sequence number, bottle position, date, time, and
beginning and ending scan numbers to a bottle log (.bl) file each time a bottle
fire confirmation is received from the water sampler. The beginning and
ending scan numbers correspond to approximately a 1.5-second duration for
each bottle. For a 911plus system, SEASAVE also automatically sets the
bottle confirm bit in the data (.dat) file for all scans within a 1.5-second
duration after a bottle firing confirmation is received from the water sampler.
To fire bottles:
1. Set up the water sampler in the Configure menu
(see Water Sampler Configuration in Section 3: Configure Menu,
Part I - General System Setup).
2. Start real-time data acquisition.
3. In the View menu, select Fire Bottle Control. The Bottle Fire dialog
box appears.
4. When desired, click Fire Bottle.
If you selected Sequential or Table driven in Water Sampler
Configuration (Step 1), the dialog box displays the number of the
fired bottle.
If you selected User Input in Water Sampler Configuration (Step 1),
SEASAVE prompts you to enter the bottle number.
Note:
The .bl file has the same file
name and is placed in the same
directory as the data file. For
example, if the data file is
test1.hex, the .bl file is test1.bl.
Note:
If desired, you can fire bottles
without using the Bottle Fire
dialog box. Each time you want to
fire a bottle, select Fire Bottle in
the RealTimeData menu, or press
Ctrl F3.