Old Style SBE 25 SEALOGGER Configuration
Listing of channels at bottom of dialog box reflects this
choice (0 - 7). Must agree with number programmed into
SBE 25 with CC command; see reply from status command
(DS). Voltage channel 0 in .con file corresponds to first
external voltage in data stream, voltage channel 1 to second
voltage in data stream, etc.
Used to determine strain gauge
pressure sensor data format. See
reply from status command (DS).
1, 2, 4, or 8 scans/second. Used to
calculate elapsed time, if you select
time as a parameter for a display
window. For elapsed time calculation to
be correct, this entry must agree with
number programmed into SBE 25 with
CC command; see reply from status
command (DS).
Surface PAR - Select if Surface
PAR voltage added by deck unit.
Selecting Surface PAR voltage
adds 2 channels to
Channel/Sensor table.
See Application Note 47.
NMEA - Select if using with a deck
unit connected to a NMEA
navigation device.SEASAVE
automatically adds current latitude,
longitude, and universal time code
to data header. Select NMEA
(Lat/Lon) Interface in SEASAVE’s
Configure menu to control how
Lat/Lon data is incorporated.
Click to modify list of sensors (after number of
frequency channels and voltage words have
been specified above). Data Format Definition
dialog box with a list of sensors appears.
Select sensors.
Click to return to Examine / Change
Instrument Configuration File dialog
box. In that dialog box, click Change
Calibration Coefficients to view,
modify, or enter coefficients.