User`s manual
Section 4: Configure Menu, Part II - Instrument .con File
38
Old Style SBE 9plus Configuration
Listing of channels at bottom of dialog
box reflects this choice. Typically:
• 0 = SBE 3 or 4 plugged into
COND 2 on SBE 9plus
end cap (dual redundant
sensor configuration)
• 1 = SBE 3 or 4 plugged into
TEMP 2 on SBE 9plus end
cap and not using COND 2
connector (single redundant sensor
configuration)
•
2 = no redundant T or C sensors
• Surface PAR - Select if Surface
PAR sensor used; must agree with
ADDSPAR= command
programmed into Deck Unit.
Adds 2 channels to
Channel/Sensor table. Do not
decrease Voltage words
suppressed to reflect this; Voltage
words suppressed reflects only
external voltages going directly to
9plus from auxiliary sensors.
• NMEA - Select if NMEA navigation
device used. If selected,
SEASAVE automatically adds
current latitude, longitude, and
universal time code to data
header. Select NMEA (Lat/Lon)
Interface in SEASAVE’s Configure
menu to control how Lat/Lon data
is incorporated.
Select to include time of
each scan with data.
Click to modify list of sensors (after number of
frequency channels and voltage words have
been specified above). Data Format Definition
dialog box with a list of sensors appears.
Select sensors.
IEEE-488 (GPIB) or
RS-232C, based on how
SBE 11plus is connected
to computer.
Listing of channels at bottom of dialog box
reflects this choice.
Total number of voltage words is 4, and each
word contains data from two 12-bit A/D
channels. SBE 11plus suppresses words
starting with highest numbered word. Number
of words to keep is determined by highest
numbered external voltage input that is not
a spare:
Words to suppress = 4 - Words to Keep
External Voltage Connector Words to Keep
(not spare)
0 or 1 AUX 1 1
2 or 3 AUX 2 2
4 or 5 AUX 3 3
6 or 7 AUX 4 4
Click to return to Examine / Change
Instrument Configuration File dialog
box. In that dialog box, click Change
Calibration Coefficients to view,
modify, or enter coefficients.