User`s manual
Section 4: Configure Menu, Part II - Instrument .con File
33
New Style SBE 45 MicroTSG Configuration
The SBE 45 transmits ASCII converted data in engineering units. It converts
the raw data internally to engineering units, based on the programmed
calibration coefficients. See the SBE 45 manual.
Define data in SBE 45 data stream:
• Output conductivity with each scan. -
Must agree with OUTPUTCOND=
command programmed into SBE 45.
• Output salinity with each scan –
Must agree with OUTPUTSAL=
command programmed into SBE 45.
• Output sound velocity with each scan
– Must agree with OUTPUTSV=
command programmed into SBE 45.
See reply from status command (DS)
for setup programmed into SBE 45.
• Use junction box - Select if SBE 45
data is transmitted to computer
through optional 90402 – SBE 45
Interface Box. Interface Box can
append optional SBE 38 and NMEA
data to SBE 45 data stream.
• SBE 38 temperature added –
Select if 90402 – SBE 45 Interface
Box is connected to SBE 38 remote
temperature sensor. Interface
Box appends SBE 38 data to
data stream.
SEASAVE (and Data Conversion
and Derive in SBE Data Processing)
uses remote temperature data
when calculating density and
sound velocity.
• NMEA data added - Select if 90402
– SBE 45 Interface Box is connected
to a NMEA navigation device.
Interface Box appends NMEA data
to data stream. SEASAVE adds
current latitude, longitude, and
universal time code to data header.
New to create new
.con file for this CTD.
Open to select
different .con file.
Save or Save As to
save current .con file
settings.
Click to exit dialog box. If you made changes and did
not Save or Save As, SEASAVE asks if you want to
save changes.
Time between scans. Must agree with number
programmed into SBE 45 with INTERVAL=
command; see reply from status command (DS).