User`s manual

Section 4: Configure Menu, Part II - Instrument .con File
24
The .con file discussion is in several parts:
Instrument Configuration - New Style (in this section):
Configuration dialog box for each instrument available in the new style
(SBE 911/917plus, 16, 16plus, 19, 19plus, 21, 25, 45, and 49).
Instrument Configuration - Old Style (in this section):
Configuration dialog box for each instrument available in the old style
(SBE 911plus, 911e, 911, 16, 19, 21, 25, and 31).
Section 5: Configure Menu, Part III - Calibration Coefficients: calculation
of calibration coefficients for each type of frequency, A/D count, and
voltage sensor.
Instrument Configuration - New Style
The discussion of new style instrument configuration is in two parts:
General description of how to view, modify, or create a .con file
Detailed description of the Configuration dialog box for each instrument
Viewing, Modifying, or Creating .con File
1. To create a new .con file: In the Configure menu, select New Style
Instrument Configuration / Create New Instrument Configuration and
select the desired instrument. Go to Step 3.
2. To select and view or modify an existing .con file:
A. In the Configure menu, select New Style Instrument Configuration /
Select Instrument Configuration. In the dialog box, browse to the
desired file and click OK.
B. In the Configure menu, select New Style Instrument Configuration /
Modify Selected Instrument Configuration.
3. The Instrument Configuration dialog box appears. The selections at the
top of the dialog box are different for each instrument. An example is
shown below for the SBE 19plus.
Note:
Unless noted otherwise, SEASAVE
supports only one of each brand and
type of auxiliary sensor (for example,
you cannot specify two Chelsea
Minitracka fluorometers, but you can
specify a Chelsea Minitracka and a
Chelsea UV Aquatracka fluorometer.
See the individual sensor
descriptions in Section 5: Configure
Menu, Part III – Calibration
Coefficients for those sensors that
SEASAVE supports in a redundant
configuration (two or more of
the same sensor interfacing with
the CTD) when using the
New Style configuration.