SEASOFT-Win32: SEASAVE CTD Real-Time Data Acquisition Software for Windows 95/98/NT/2000/XP User’s Manual Sea-Bird Electronics, Inc. 1808 136th Place NE Bellevue, Washington 98005 USA Telephone: 425-643-9866 Fax: 425-643-9954 E-mail: seabird@seabird.com Website: www.seabird.com 10/24/05 Software Release 5.
Limited Liability Statement Extreme care should be exercised when using or servicing this equipment. It should be used or serviced only by personnel with knowledge of and training in the use and maintenance of oceanographic electronic equipment. SEA-BIRD ELECTRONICS, INC. disclaims all product liability risks arising from the use or servicing of this system. SEA-BIRD ELECTRONICS, INC.
Table of Contents Table of Contents Section 1: Introduction ....................................................................................5 How to Contact Sea-Bird ...................................................................................5 Summary............................................................................................................5 System Requirements .................................................................................6 Products Supported..................
Table of Contents Section 5: Configure Menu, Part III - Calibration Coefficients ................46 Accessing Calibration Coefficients Dialog Boxes ...........................................46 Calibration Coefficients for Frequency Sensors ..............................................47 Temperature Calibration Coefficients.......................................................47 Conductivity Calibration Coefficients ......................................................
Section 1: Introduction Section 1: Introduction This section includes contact information and a brief description of SEASOFT-Win32 and its components. How to Contact Sea-Bird Sea-Bird Electronics, Inc. 1808 136th Place Northeast Bellevue, Washington 98005 USA Telephone: E-mail: 425-643-9866 seabird@seabird.com Fax: Website: 425-643-9954 http://www.seabird.
Section 1: Introduction System Requirements Sea-Bird recommends the following minimum system requirements for SEASOFT-Win32: Pentium 90 CPU, 64 Mbyte RAM, Windows 98 or later.
Section 2: Installation and Use Section 2: Installation and Use SEASAVE requires approximately 9 Mbytes of disk space during installation. Ensure there is room on your hard drive before proceeding. Sea-Bird recommends the following minimum system requirements for SEASOFT-Win32: Pentium 90 CPU, 64 Mbyte RAM, Windows 98 or later. Installation 1. Note: Sea-Bird supplies the current version of our software when you purchase an instrument.
Section 2: Installation and Use SEASAVE Use SEASAVE Window Note: SEASAVE can be run from the command line. See Appendix I: Command Line Operation. To start SEASAVE: • Double click on seasave.
Section 2: Installation and Use • Notes: • The algorithms used to calculate derived parameters in SEASAVE are the same as used in SBE Data Processing’s Derive and Data Conversion modules (with the exception of the oxygen calculation). See the SBE Data Processing manual on our website or the SBE Data Processing Help files for details.
Section 2: Installation and Use Getting Started Displaying Archived Data - Sea-Bird Demo Files SEASAVE can be used to display archived raw data in a .hex or .dat file. Sea-Bird provides example files with the software to assist you in learning how to use SEASAVE. These files are automatically installed on your hard drive when you install SEASAVE; the default location is: C:\Program Files\Sea-Bird\SS-WIN32-DEMO The demo files include: • one data file - demo.dat • one instrument configuration file - demo.
Section 2: Installation and Use Acquiring and Displaying Real-Time Data Follow these steps to get started using SEASAVE to acquire and display real-time data: 1. Set up the instrument, and define SEASAVE operating parameters (see Sections 3, 4, and 5: Configure Menu): • Instrument Configuration (.con) File - define what sensors are integrated with the instrument, each sensor’s calibration coefficients, and what other data is integrated with the data stream from the instrument.
Section 2: Installation and Use File Formats File extensions are used by SEASOFT to indicate the file type. Input files for real-time data acquisition: Extension .con .cfg .dsf .dso .dss Description Instrument configuration - number and type of sensors, channel assigned to each sensor, and calibration coefficients. SEASAVE uses this information to interpret the raw data from the instrument. Latest version of .
Section 2: Installation and Use Output files from real-time data acquisition: Extension .bl .dat .hex .hdr .mrk .nav .txt Description Bottle log information - output bottle file, containing bottle firing sequence number and position, date, time, and beginning and ending scan numbers for each bottle closure. Beginning and ending scan numbers correspond to approximately a 1.5-second duration for each bottle.
Section 3: Configure Menu, Part I - General System Setup Section 3: Configure Menu, Part I - General System Setup Note: Setup of all parameters in the Configure menu is included in the SEASAVE configuration (.cfg) file. To save the setup, you must save the .cfg file (File menu / Save Seasave configuration as . . .) before exiting SEASAVE.
Section 3: Configure Menu, Part I - General System Setup Remote Display Alarm Note: Remote Display alarm setup in SEASAVE is applicable only for an SBE 14 connected to a computer COM port. If the SBE 14 is connected directly to an SBE 11plus Deck Unit, Remote Display alarm setup is done by sending commands to the SBE 11plus using SEATERM. The SBE 14 Remote Display can be set up to turn an alarm on when pressure is less than and/or more than a specified amount.
Section 3: Configure Menu, Part I - General System Setup ASCII Output SEASAVE can output ASCII data (converted data in engineering units) to a COM port on your computer or to a shared file on your computer. If outputting ASCII data to a shared file, you can use Word, Notepad, or some other program to open and look at the data while SEASAVE continues to acquire more data.
Section 3: Configure Menu, Part I - General System Setup Header Form Note: A header is automatically included in the data (.dat or .hex) file and in the header (.hdr) file. The header includes software version, sensor serial numbers, instrument configuration, date and time of start of data acquisition, etc. There can be up to two date/time listings in the header. The first, System Upload Time, is always the date and time from the computer.
Section 3: Configure Menu, Part I - General System Setup When you begin data acquisition, (if you chose to store the data on disk) the header form appears for you to fill in. The user-selected prompts (Ship, Cruise, Station, Latitude, and Longitude) appear to the left of the blank fields.
Section 3: Configure Menu, Part I - General System Setup Mark Variable Selection Note: The .mrk file has the same file name as the data file. For example, if the data file is test1.hex, the .mrk file is test1.mrk. Mark Scan allows you to copy the most recent scan of data to a mark (.mrk) file as desired. The .
Section 3: Configure Menu, Part I - General System Setup NMEA Lat/Lon Interface Notes: • There can be up to two date/time listings in the header. The first, System Upload Time, is always the date and time from the computer. The second, UTC Time, is the date and time from an optional NMEA navigation device. • NMEA Lat/Lon Interface selections are not applicable to the SBE 45. The 90402 – SBE 45 Interface Box defines what navigation data is included in the data file.
Section 3: Configure Menu, Part I - General System Setup Remote Display The SBE 14 Remote Display can display depth, pressure, and/or altimeter height for a CTD system. To enable and set up the Remote Display: 1. In the Configure menu, select Remote Display. The Remote Display Configuration dialog box appears: Must be selected to enable sending data to SBE 14, and to enable Remote Display Alarm (see Alarms above).
Section 3: Configure Menu, Part I - General System Setup Water Sampler Configuration For real-time data acquisition, a Sea-Bird CTD can be integrated with a water sampler when used with a deck unit. The water sampler bottles can be fired by command from SEASAVE (see Firing Bottles in Section 7: Real-Time Data Acquisition).
Section 4: Configure Menu, Part II - Instrument .con File Section 4: Configure Menu, Part II Instrument .con File Note: Setup of all parameters in the Configure menu, including the name and location of the selected .con file, is included in the SEASAVE configuration (.cfg) file. To save the setup, you must save the .cfg file (File menu / Save Seasave configuration as . . .) before exiting SEASAVE. This section describes the setup of the instrument configuration (.con) file in the Configure menu.
Section 4: Configure Menu, Part II - Instrument .con File The .con file discussion is in several parts: • Instrument Configuration - New Style (in this section): Configuration dialog box for each instrument available in the new style (SBE 911/917plus, 16, 16plus, 19, 19plus, 21, 25, 45, and 49). • Instrument Configuration - Old Style (in this section): Configuration dialog box for each instrument available in the old style (SBE 911plus, 911e, 911, 16, 19, 21, 25, and 31).
Section 4: Configure Menu, Part II - Instrument .con File All Instrument Configuration dialog boxes include: • List of instrument configuration options at the top (instrumentspecific), such as number of auxiliary channels, pressure sensor type, addition of Surface PAR and NMEA to the CTD data string. • Channel/Sensor Table: This table reflects the options selected at the top (for example, the number of voltage sensors listed in the table agrees with the user-selection for External voltage channels).
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 9plus Configuration Channel/Sensor table reflects this choice. Voltage channel 0 in .con file corresponds to sensor wired to channel 0 on end cap connector, voltage channel 1 to sensor wired to channel 1 on end cap connector, etc. Total number of voltage words is 4; each word contains data from two 12-bit A/D channels. Deck Unit and SEARAM suppress words above highest numbered voltage word used.
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 16 SEACAT C-T Recorder Configuration Channel/Sensor table reflects this choice. Must agree with number programmed into SBE 16 with SVn (n=0, 1, 2, 3, or 4) command; see reply from status command (DS). Voltage channel 0 in .con file corresponds to sensor wired to channel 0 on end cap connector, voltage channel 1 in .con file corresponds to sensor wired to channel 1 on end cap connector, etc.
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 16plus SEACAT C-T Recorder Configuration The SBE 16plus can interface with one SBE 38 secondary temperature sensor, one SBE 50 pressure sensor, or up to two Pro-Oceanus Gas Tension Devices (GTDs) through the SBE 16plus optional RS-232 connector. Data from an SBE 50 pressure sensor is appended to the data stream, and does not replace the (optional) internally mounted pressure sensor data.
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 19 SEACAT Profiler Configuration SEASAVE always treats the SBE 19 as if it is a Profiling instrument (i.e., it is in Profiling mode). If your SBE 19 is in Moored Mode, you must treat it like an SBE 16 (when setting up the .con file, select the SBE 16). Select strain gauge or Digiquartz with temperature compensation. Channel/Sensor table reflects this choice.
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 19plus SEACAT Profiler Configuration Channel/Sensor table reflects this choice (0, 1, 2, 3, or 4). Must agree with number programmed into 19plus with VOLTn= commands (n= 0, 1, 2, and 3); see reply from status command (DS). Voltage channel 0 in .con file corresponds to first external voltage in data stream, voltage channel 1 to second external voltage in data stream, etc. Select strain gauge (only selection applicable to 19plus).
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 21 Thermosalinograph Configuration Channel/Sensor table reflects this choice (shows additional frequency-based temperature channel if SBE 3 selected, or RS-232 channel if SBE 38 selected). Must agree with SBE3= or SBE38= command programmed into SBE 21 to enable or disable external temperature sensor; see reply from status command (DS).
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 25 SEALOGGER Configuration Channel/Sensor table reflects this choice (0 - 7). Must agree with number programmed into SBE 25 with CC command; see reply from status command (DS). Voltage channel 0 in .con file corresponds to first external voltage in data stream, voltage channel 1 to second external voltage in data stream, etc. 1, 2, 4, or 8 scans/second.
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 45 MicroTSG Configuration The SBE 45 transmits ASCII converted data in engineering units. It converts the raw data internally to engineering units, based on the programmed calibration coefficients. See the SBE 45 manual. Time between scans. Must agree with number programmed into SBE 45 with INTERVAL= command; see reply from status command (DS). Define data in SBE 45 data stream: • Output conductivity with each scan.
Section 4: Configure Menu, Part II - Instrument .con File New Style SBE 49 FastCAT Configuration Number of samples to average per scan. SBE 49 samples at 16 Hz (0.0625 seconds), averages data, and transmits averaged data realtime. Must agree with number programmed into SBE 49 with NAVG= command; see reply from status command (DS). New to create new .con file for this CTD. Open to select different .con file. Save or Save As to save current .con file settings.
Section 4: Configure Menu, Part II - Instrument .con File Instrument Configuration - Old Style The discussion of old style instrument configuration is in two parts: • • General description of how to view, modify, or create a .con file Detailed description of the Configuration dialog box for each instrument Viewing, Modifying, or Creating .con File Note: Unless noted otherwise, SEASAVE supports only one of each brand and type of auxiliary sensor interfacing with a CTD.
Section 4: Configure Menu, Part II - Instrument .con File B. Click Change Instrument Configuration to change or set up the configuration of the instrument (number and type of sensors, channel assigned to each sensor, etc.). The Instrument Configuration dialog box appears. This dialog box is different for each instrument. An example is shown below for the SBE 19.
Section 4: Configure Menu, Part II - Instrument .con File C. After completing instrument configuration, click Change Calibration Coefficients (in the Examine / Change Instrument Configuration File dialog box) to change or enter sensors’ serial numbers, calibration dates, and/or calibration coefficients. The Change Calibration Coefficients dialog box appears, with a list of the sensors you defined for the configuration in Step 4B: Double click on a sensor to bring up its calibration coefficient dialog box.
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 9plus Configuration Listing of channels at bottom of dialog box reflects this choice. Total number of voltage words is 4, and each word contains data from two 12-bit A/D channels. SBE 11plus suppresses words starting with highest numbered word.
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 911 Configuration Select from standard or high resolution Digiquartz, with or without temperature compensation, or strain gauge. If pressure sensor calibration sheet contains only C, D, and T0 coefficients, you do not have temperature compensation. If pressure sensor calibration sheet contains C1, C2, C3, D1, T1, T2, and T3 coefficients, you do have temperature compensation.
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 911e Configuration Select from standard or high resolution Digiquartz, with or without temperature compensation, or strain gauge. If pressure sensor calibration sheet contains only C, D, and T0 coefficients, you do not have temperature compensation. If pressure sensor calibration sheet contains C1, C2, C3, D1, T1, T2, and T3 coefficients, you do have temperature compensation.
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 16 SEACAT C-T Recorder Configuration Listing of channels at bottom of dialog box reflects this choice (0, 1, or 2); typically 0. Must agree with setup of SBE 16; external frequency channels were a custom application for the SBE 16 see configuration sheet. Select strain gauge, Digiquartz with or without temperature compensation, or no pressure sensor.
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 19 SEACAT Profiler Configuration SEASAVE always treats the SBE 19 as if it is a Profiling instrument (i.e., it is in Profiling mode). If your SBE 19 is in Moored Mode, you must treat it like an SBE 16 (when setting up the .con file, select the SBE 16). Listing of channels at bottom of dialog box reflects this choice. Must agree with number programmed into SBE 19 with SVn (n=0, 2, or 4) command; see reply from status command (DS).
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 21 Thermosalinograph Configuration Listing of channels at bottom of dialog box reflects this choice (0 or 1). Must agree with XY or XN command programmed into SBE 21 to enable or disable external frequency (temperature) sensor; see reply from status command (DS). If external frequency is enabled, SEASAVE (and Data Conversion and Derive in SBE Data Processing) uses remote temperature data when calculating density and sound velocity.
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 25 SEALOGGER Configuration Listing of channels at bottom of dialog box reflects this choice (0 - 7). Must agree with number programmed into SBE 25 with CC command; see reply from status command (DS). Voltage channel 0 in .con file corresponds to first external voltage in data stream, voltage channel 1 to second voltage in data stream, etc. Used to determine strain gauge pressure sensor data format. See reply from status command (DS).
Section 4: Configure Menu, Part II - Instrument .con File Old Style SBE 31 Configuration Most SBE 31 CTD systems do not have a pressure sensor. Listing of channels at bottom of dialog box reflects these choices. Select if NMEA interface option installed in SBE 31 and SBE 31 is connected to a NMEA navigation device. SEASAVE automatically adds current latitude, longitude, and universal time code to data header.
Section 5: Configure Menu, Part III - Calibration Coefficients Section 5: Configure Menu, Part III Calibration Coefficients This section describes the calculation and/or source of the calibration coefficients for the configuration (.con) file, for each type of sensor supported by Sea-Bird CTDs. SEASAVE uses the sensor calibration coefficients to convert the raw data stream into engineering units for display during real-time data acquisition.
Section 5: Configure Menu, Part III - Calibration Coefficients Calibration Coefficients for Frequency Sensors For all calibration dialog boxes, enter the sensor serial number and calibration date. Many sensor calibration equations contain an offset term. Unless noted otherwise, use the offset (default = 0.0) to make small corrections for sensor drift between calibrations. Calibration coefficients are discussed below for each type of sensor.
Section 5: Configure Menu, Part III - Calibration Coefficients Note: Use coefficients g, h, i, j, Ctcor, and Cpcor (if available on calibration sheet) for most accurate results; conductivity for older sensors was calculated based on a, b, c, d, m, and Cpcor. Conductivity Calibration Coefficients Enter g, h, i, j, Ctcor (or a, b, c, d, m) and Cpcor from the calibration sheet. • Cpcor makes a correction for the highly consistent change in dimensions of the conductivity cell under pressure.
Section 5: Configure Menu, Part III - Calibration Coefficients Pressure (Paroscientific Digiquartz) Calibration Coefficients Note: See Calibration Coefficients for A/D Count Sensors below for information on strain gauge pressure sensors used on the SBE 16plus, 19plus, and 49. See Calibration Coefficients for Voltage Sensors below for information on strain gauge pressure sensors used on other instruments. Enter the sets of C, D, and T coefficients from the calibration sheet.
Section 5: Configure Menu, Part III - Calibration Coefficients Calibration Coefficients for A/D Count Sensors For all calibration dialog boxes, enter the sensor serial number and calibration date. Many sensor calibration equations contain an offset term. Unless noted otherwise, use the offset (default = 0.0) to make small corrections for sensor drift between calibrations. Calibration coefficients are discussed below for each type of sensor: temperature and strain gauge pressure sensor.
Section 5: Configure Menu, Part III - Calibration Coefficients Calibration Coefficients for Voltage Sensors For all calibration dialog boxes, enter the sensor serial number and calibration date. Many sensor calibration equations contain an offset term. Unless noted otherwise, use the offset (default = 0.0) to make small corrections for sensor drift between calibrations. Calibration coefficients are discussed below for each type of sensor.
Section 5: Configure Menu, Part III - Calibration Coefficients • Note: See Application Note 39 for complete description of calculation of Chelsea Aqua 3 calibration coefficients. Chelsea Aqua 3 Enter VB, V1, Vacetone, slope, offset, and SF. Concentration (µg/l) = slope*[(10.0(V/SF) - 10.0VB)/(10.0V1 - 10.0Vacetone)] + offset where VB, V1, and Vacetone are from calibration sheet Slope (default 1.0) and offset (default 0.0) adjust readings to conform to measured concentrations Scale factor SF = 1.
Section 5: Configure Menu, Part III - Calibration Coefficients • Seapoint Enter gain and offset. Concentration = (V * 30/gain) + offset where Gain is dependent on cable used (see cable drawing, pins 5 and 6) Note: SEASAVE can process data for an instrument interfacing with up to two Seapoint fluorometers when using the New Style configuration. • Seapoint Rhodamine (New Style configuration only) Enter gain and offset.
Section 5: Configure Menu, Part III - Calibration Coefficients • Turner SCUFA (New Style configuration only) Enter scale factor, offset, units, mx, my, and b from the calibration sheet.
Section 5: Configure Menu, Part III - Calibration Coefficients Methane Sensor Calibration Coefficients (New Style configuration only) The Capsum METS sensor requires two channels – one for the methane concentration and the other for the temperature measured by the sensor. Make sure to select both when configuring the instrument. For the concentration channel, enter D, A0, A1, B0, B1, and B2.
Section 5: Configure Menu, Part III - Calibration Coefficients • IFREMER This sensor requires two channels - one for the direct voltage and the other for the measured voltage. Make sure to select both when configuring the instrument. For the direct voltage channel, enter vm0, vd0, d0, and k.
Section 5: Configure Menu, Part III - Calibration Coefficients Oxygen Calibration Coefficients Notes: • Enter soc and boc values from the most recent field calibration for Beckman-type, YSI-type, or SeaBird (SBE 43) oxygen sensor. • See Application Notes 13-1 and 13-3 for complete description of calculation of calibration coefficients for Beckman- or YSI-type sensors. • See Application Notes 64 and 64-2 for complete description of calculation of SBE 43 calibration coefficients.
Section 5: Configure Menu, Part III - Calibration Coefficients • Notes: • Selection of Par / Irradiance, Biospherical / Licor as the voltage sensor is also applicable to the Chelsea PAR sensor. • For complete description of calculation of surface PAR calibration coefficients, see Application Note 11S (SBE 11plus Deck Unit) or 47 (SBE 33 or 36 Deck Unit). Chelsea PAR sensor Calibration constant = 10 9 / 0.01 (for units of microEinsteins/sec–m2) or = 10 9 / 0.04234 (for units of quanta/sec–m2) M = 1.
Section 5: Configure Menu, Part III - Calibration Coefficients Transmissometer Calibration Coefficients Note: See Application Note 7 for complete description of computation of M and B. • Sea Tech, Chelsea (Alphatracka), and WET Labs Cstar Enter M, B, and path length (in meters) Path length (distance between lenses) is based on sensor size (for example, 25 cm transmissometer = 0.25m path length, etc.).
Section 5: Configure Menu, Part III - Calibration Coefficients User Polynomial (for user-defined sensor) Calibration Coefficients The user polynomial allows you to define an equation to relate the sensor output voltage to calculated engineering units, if your sensor is not pre-defined in Sea-Bird software. Enter a0, a1, a2, and a3. Val = a0 + (a1 * V) + (a2 * V2) + (a3 * V3) where: V = voltage from sensor a0, a1, a2, and a3 = user-defined sensor polynomial coefficients If desired, enter the sensor name.
Section 6: ScreenDisplay Menu - Setting Up SEASAVE Displays Section 6: ScreenDisplay Menu Setting Up SEASAVE Displays This section describes how to set up and arrange SEASAVE display windows. SEASAVE can have up to ten display windows. Edit a display window to select desired parameters, number of digits for data display, and plot characteristics (labels, grids, etc.). This information is saved in a setup file for each display window. File extensions vary, depending on display type: .
Section 6: ScreenDisplay Menu - Setting Up SEASAVE Displays Fixed Display or Scrolled Display Setup for the Fixed and Scrolled Displays are similar. The Fixed Display Setup dialog box looks like this: Click on Modify Display Parameters to get the Fixed Display Variables Setup dialog box: Select the desired variable for each row by clicking Select Variable. A dialog box with a list of variables appears; make your selection and click OK.
Section 6: ScreenDisplay Menu - Setting Up SEASAVE Displays Overlay Plot Display The Display Setup dialog box looks like this: The dialog box selections and buttons include: • Plot Label: Label placed at top, center of plot. • Show Fire Sequence, Show Bottle Lines, and Bottle Line Configuration: For a system integrated with a water sampler. If Show Fire Sequence selected, SEASAVE lists the bottle closure order to the right of the plot.
Section 6: ScreenDisplay Menu - Setting Up SEASAVE Displays Click on Modify Display Parameters to get the Overlay Display Parameters Setup dialog box: The dialog box entries include: • Grid and Colors: Select a background color for the plot, and select whether to show a grid, along with the grid style and color. • Variable selections and plotting parameters: SEASAVE can plot one parameter on the y-axis and up to four parameters on the x-axis.
Section 7: Real-Time Data Acquisition Section 7: Real-Time Data Acquisition Note: To start acquisition without a mouse: • Windows 2000 / XP – Press the Alt key to show the keyboard shortcuts (underlines) on menus. Press the appropriate letter (for example, R for RealtimeData menu) and use the arrow and Enter keys to navigate. • Windows 95 / 98 / NT – Keyboard shortcuts (underlines) appear on menus at all times. Press the Alt key and appropriate letter and use the arrow and Enter keys to navigate.
Section 7: Real-Time Data Acquisition • COMM Port Configuration: Click to configure the transfer of data. The COMM Port Configuration dialog box varies, depending on your system setup. ! SBE 911, 911e, or 911plus, with or without water sampler, the dialog box looks like this: Computer port connected to deck unit SBE 11 Interface connector; sends commands to and receives replies from CTD Baud rate between deck unit and computer; must agree with deck unit setting.
Section 7: Real-Time Data Acquisition • Start Acquire: Begin processing and displaying data. A. If you selected Store on Disk above, and selected Prompt for Header Information in the Header Form setup (Configure menu), the Header Information dialog box appears. Fill in the desired header and click OK. B.
Section 7: Real-Time Data Acquisition Firing Bottles Note: The .bl file has the same file name and is placed in the same directory as the data file. For example, if the data file is test1.hex, the .bl file is test1.bl. Water sampler bottles can be fired by command from SEASAVE. SEASAVE automatically writes bottle sequence number, bottle position, date, time, and beginning and ending scan numbers to a bottle log (.bl) file each time a bottle fire confirmation is received from the water sampler.
Section 7: Real-Time Data Acquisition Marking Scans Note: The .mrk file has the same file name and is placed in the same directory as the data file. For example, if the data file is test1.hex, the .mrk file is test1.mrk. Mark Scan allows you to copy the most recent scan of data to a mark (.mrk) file as desired. The .
Section 7: Real-Time Data Acquisition Turning Pump On / Off SEASAVE allows you to manually turn a 911plus’ pump on and off during data acquisition. This may be useful if your system is integrated with an acoustic instrument, to provide a quiet period during its data acquisition. To manually turn SBE 911plus pump on / off: Note: You must start SEASAVE from the command line, using the -pc command, to enable pump turn on / off from SEASAVE.
Section 8: Displaying Archived Data Section 8: Displaying Archived Data Note: To display archived data without a mouse: • Windows 2000 / XP – Press the Alt key to show the keyboard shortcuts (underlines) on menus. Press the appropriate letter (for example, A for ArchivedData menu) and use the arrow and Enter keys to navigate. • Windows 95, 98, and NT – Keyboard shortcuts (underlines) appear on menus at all times. Press the Alt key and appropriate letter and use the arrow and Enter keys to navigate.
Section 8: Displaying Archived Data 2. To pause and restart data display: In the ArchivedData menu, select Pause. The data display stops, but SEASAVE retains information on where it stopped. In the ArchivedData menu, select Continue when ready to restart the display where it stopped. 3. To adjust rate that data is displayed: In the ArchivedData menu, select Faster, Slower, or No Wait. No Wait plays back data at the rate at which it was acquired. 4.
Section 9: Processing Data Section 9: Processing Data Sea-Bird provides software, SBE Data Processing, for converting the raw .hex or .dat data file into engineering units, editing (aligning, filtering, removing bad data, etc.) the data, calculating derived variables, and plotting the processed data. However, sometimes users want to edit the raw .hex or .
Section 9: Processing Data 5. In the File menu, select Save (not Save As). If you are running Windows 2000, the following message displays: You are about to save the document in a Text-Only format, which will remove all formatting. Are you sure you want to do this? Ignore the message and click Yes. 6. In the File menu, select Exit. .dat Files Sea-Bird is not aware of a technique for editing a .dat file that will not corrupt it. Opening a .
Appendix I: Command Line Operation Appendix I: Command Line Operation SEASAVE has several command line parameters, for infrequently used options: Parameter Function -autostart= Automatically start SEASAVE and data acquisition, saving data filename to filename. Filename must include path and extension (.dat for SBE 9plus; .hex for SBE 16, 16plus, 19, 19plus, 21, 25, or 31). SEASAVE uses .con file and display setup from last saved SEASAVE configuration (.cfg).
Appendix I: Command Line Operation To run SEASAVE with a Command Line Parameter: Notes: • If the path includes any spaces, enclose the path in quotes (“path“). See the examples. • An alternative method of running SEASAVE with a Command Line Parameter is from a command prompt. 1. In the Windows Start menu, select Run. The Run dialog box appears. Enter the command line parameter(s) as shown below: Path\seasave.exe parameter1 parameter2 . . . where Path is the location of seasave.
Appendix II: Configure (.con) File Format Appendix II: Configure (.con) File Format Shown below is a line-by-line description of the .con file contents, which can be viewed in a text editor.
Appendix II: Configure (.
Appendix II: Configure (.
Appendix II: Configure (.
Appendix III: Software Problems Appendix III: Software Problems Considerable effort has been made to test and check this software before its release. However, because of the wide range of instruments that Sea-Bird produces (and interfaces with) and the many applications that these instruments are used in, there may be software problems that have not been discovered and corrected. If a problem occurs, please contact us via phone (425-643-9866), email (seabird@seabird.
Appendix IV: Derived Parameter Formulas Appendix IV: Derived Parameter Formulas Note: Algorithms used for calculation of derived parameters in SEASAVE and in SBE Data Processing’s Data Conversion, Derive, and SeacalcW modules are identical, except as noted. For formulas for the calculation of conductivity, temperature, and pressure, see the calibration sheets for your instrument.
Appendix IV: Derived Parameter Formulas density = ρ = ρ (s, t, p) [kg/m3] (density of seawater with salinity s, temperature t, and pressure p, based on the equation of state for seawater (EOS80)) Density calculation: Using the following constants B0 = 8.24493e-1, B1 = -4.0899e-3, B2 = 7.6438e-5, B3 = -8.2467e-7, B4 = 5.3875e-9, C0 = -5.72466e-3, C1 = 1.0227e-4, C2 = -1.6546e-6, D0 = 4.8314e-4, A0 = 999.842594, A1 = 6.793952e-2, A2 = -9.095290e-3, A3 = 1.001685e-4, A4 = -1.120083e-6, A5 = 6.
Appendix IV: Derived Parameter Formulas depth = [m] Depth calculation: C Computer Code – // Depth double Depth(int dtype, double p, double latitude) // dtype = fresh water or salt water, p = pressure in decibars, latitude in degrees { double x, d, gr; if (dtype == FRESH_WATER) /* fresh water */ d = p * 1.019716; else { /* salt water */ x = sin(latitude / 57.29578); x = x * x; gr = 9.780318 * (1.0 + (5.2788e-3 + 2.36e-5 * x) * x) + 1.092e-6 * p; d = (((-1.82e-15 * p + 2.279e-10) * p - 2.2512e-5) * p + 9.
Appendix IV: Derived Parameter Formulas sound velocity = [m/sec] (sound velocity can be calculated as Chen-Millero, DelGrosso, or Wilson) Sound velocity calculation: C Computer Code – // Sound Velocity Chen and Millero double SndVelC(double s, double t, double p0) /* sound velocity Chen and Millero 1977 */ /* JASA,62,1129-1135 */ // s = salinity, t = temperature deg C ITPS-68, p = pressure in decibars { double a, a0, a1, a2, a3; double b, b0, b1; double c, c0, c1, c2, c3; double p, sr, d, sv; p = p0 / 10.
Appendix IV: Derived Parameter Formulas p=p Σ average sound velocity = ∆p,p=min di [m/s] p=p Σ ∆p,p=min di / vi Average sound velocity is the harmonic mean (average) from the surface to the current CTD depth. The average is calculated on the downcast only. The first window begins when pressure is greater than a minimum specified pressure and salinity is greater than a minimum specified salinity. Depth is calculated from pressure based on user-input latitude.
Appendix IV: Derived Parameter Formulas potential temperature [IPTS-68] = θ (s, t, p, pr) [° C] (Potential temperature is the temperature an element of seawater would have if raised adiabatically with no change in salinity to reference pressure pr. Sea-Bird software uses a reference pressure of 0 decibars).
Appendix IV: Derived Parameter Formulas Descent rate and acceleration computed by SEASAVE and SBE Data Processing’s Data Conversion module are somewhat different from values computed by SBE Data Processing’s Derive module, because the algorithms calculate the derivative of the pressure signal with respect to time, using a linear regression to determine the slope.
Index Index Configure calibration coefficients · 46 calibration coefficients – A/D count sensors · 50 calibration coefficients - frequency sensors · 47 calibration coefficients - voltage sensors · 51 SBE 16 · 27, 41 SBE 16plus · 28 SBE 19 · 29, 42 SBE 19plus · 30 SBE 21 · 31, 43 SBE 25 · 32, 44 SBE 31 · 45 SBE 45 · 33 SBE 49 · 34 SBE 911 · 39 SBE 911e · 40 SBE 911plus · 26, 38 Configure menu · 14 . .
Index M S Mark scan · 19, 69 Methane · 55 SBE 16 · 27, 41 SBE 16plus · 28 SBE 19 · 29, 42 SBE 19plus · 30 SBE 21 · 31, 43 SBE 25 · 32, 44 SBE 31 · 45 SBE 32 · 22, 68 SBE 45 · 33 SBE 49 · 34 SBE 911 · 39 SBE 911e · 40 SBE 911plus · 26, 38 SBE 9plus · 38 Scrolled display · 62 Sea-Bird · 5 SEACON · 23, 77 SEASAVE file extensions · 12 file formats · 12 getting started · 10 problems · 81 toolbar · 8 use · 8 window · 8 SEASOFT file extensions · 12 file formats · 12 SEASOFT-DOS · 6 SEASOFT-Win32 installation ·