User's Manual
Indication
Error description
Error: Expired strip
The test strips are expired.
Solution
Discard this test strip and test again using a new test strip
does not past expiration date.
Indication
Error description
Error: Temperature Error Unknown barcode
If the environmental temperature is above or below the
operating range of a analyzer, a thermo analyzer icon will
appear on the display.
Solution
Move to an area at 68-88°F (18-32°C), and perform a
test. Do not artificially heat or cool the analyzer.
Indication
Error description
Error: Communication Error
The communication between analyzer and computer or
printer is failed.
Solution
Connect again between analyzer and external device.
If the error continues turn off an analyzer. Then turn on
the analyzer again. If there is still error massage, please
contact STANDARD DIAGNOSTICS, INC. TEL: 1-877-
882-3695.