User's Manual

Performing a measurement
1. Prepare a test strip for the required measurement. Check the expiry date of the test
strip. Always use test strips before their expiry date has passed.
2. Hold the test strip with thumb and index finger so that blood apply part is facing
upwards.
3. Insert the test strip into the HbA1c Test Strip Slot of the analyzer. When the test
strip reaches the correct position, Strip is automatically inserted deeper to the
analyzer. If the strip is normal for test, the blood and START button icon displays
This is blood waiting state.
Check the following displays before performing the measurement:
1 Are date and time correct?
2 Does the battery symbol or DC plug symbol appear?
3 With HbA1c test strips, check the reaction area for discolouration before starting
measurement. If you detect discolouration, this test strip is unusable. For detailed
information, see the insert of the respective test strip.