Indication
Description
Codechip Error
A code chip of other
manfuacutrers or
other products made by SD
Biosensor is inserted.
Solution Check to ensure that the code chip is
from the package of the test strip inserted.
Insert the correct one.
Flap Open Error The door of the analyzer is
open during the measurement.
Solution Close the door and do not touch or
open the door during the measurement.
Blood Error The door
is closed for testing
without a blood sample.
Solution Re-test with a blood sample.
Communication Error with Code Chip The
analyzer cannot communicate with the inserted
code chip.
Solution Re-insert the code chip. If the error
message still appears, please contact our
customer service center at 1-855-486-2888.
Internal Error An i
nternal error of the
analyzer has occurred.
Solution Turn off the analyzer, and then turn it
on again. If the error message still appears,
please contact our customer service center at
1-855-486-2888.