User's Manual
Table Of Contents
- CHAPTER 1. Understanding THE New Analyzer
- CHAPTER 2. Using the SD LipidoCare Professional Analyzer
- CHAPTER 3. Collecting blood samples
- CHAPTER 4. Performing a test for lipid or glucose
- CHAPTER 5. Control Solution Test
- CHAPTER 6. CHECK strip Test
- CHAPTER 7. Maintenance and Troubleshooting
- CHAPTER 8. Screen Message and Troubleshooting
- CHAPTER 9. Warnings, Precautions, and Limitations
- CHAPTER 10. Technical Information
- Appendix 1: References
- Appendix 2: Warning
3. Procedure for the glucose control solution
1) Take a new glucose test strip from the test strip container. Be sure to tightly replace the
container cap after taking a glucose test strip.
2) Insert a glucose test strip (yellow window with a printed arrow symbol facing up) into the
glucose test strip slot. The analyzer turns on automatically.
3) Press the left button and hold it for 3 seconds, then check to ensure that the control
solution icon appears on the screen. If you don’t want to perform a control solution test,
press the left or right button one time to exit.
4) Shake the SD Glucose Control Solution bottle and gently squeeze the container to form
one small drop. Bring the drop to the edge of the glucose strip and allow the glucose strip
to automatically draw the SD Glucose Control Solution into the yellow window. When the
SD Glucose Control Solution is applied to the glucose test strip, the Analyzer counts down
from 5 to 1 (seconds) on the screen.
5) The SD Glucose Control Solution result appears on the screen in just 5 seconds.
6) Compare the glucose control solution result to the range printed on the glucose test strip
container. If the results are not within the glucose control range printed on the glucose
test strip container, then the Analyzer and glucose strips may not be working properly.
Repeat the SD Glucose Control Solution test. If the control solution result is still not
acceptable, please contact our customer service center at 1-855-486-2888.
7) Remove the used glucose test strip from the analyzer and discard it.
CAUTION
• The acceptable range printed on the glucose test strip container is for the SD Glucose
Control Solution only. It is not a recommended range for your blood glucose level.
• Store the SD Glucose Control Solution, after closing the cap tightly,
at temperatures
between 46–86°F (8–30°C). Do not refrigerate.
• The SD Glucose Control Solution should be tested at room temperature, 64–86°F (10–
90%
RH).
• If a glucose control solution test is still not acceptable, please contact
our customer
service center at 1-855-486-2888.
- 42 -