User's Manual
Table Of Contents
- CHAPTER 1. Understanding THE New Analyzer
- CHAPTER 2. Using the SD LipidoCare Professional Analyzer
- CHAPTER 3. Collecting blood samples
- CHAPTER 4. Performing a test for lipid or glucose
- CHAPTER 5. Control Solution Test
- CHAPTER 6. CHECK strip Test
- CHAPTER 7. Maintenance and Troubleshooting
- CHAPTER 8. Screen Message and Troubleshooting
- CHAPTER 9. Warnings, Precautions, and Limitations
- CHAPTER 10. Technical Information
- Appendix 1: References
- Appendix 2: Warning
CHAPTER 5. CONTROL SOLUTION TEST
1. Control Solution Test
It is important to perform a control solution test with more than one level of the control
solution. The control solution is used to check that the analyzer and the test strips are working
together, as a system, properly. It is very important that you do this simple check routinely to
make sure you get an accurate result.
When to use Lipid or Glucose Control Solution
• Before using your analyzer for the first time
• When you open a new strip vial, pouch, or packaging
• Whenever your result does not agree with your symptoms
CAUTION
• Use only the specified control solution.
SD Lipid Control Solution
SD Glucose Control Solution
• Check the expiration date on the control solution container. Do not use the control solution
after the expiration or discard date.
• Do not swallow the control solution; it is not for human consumption.
•
Do not apply the control solution to the skin or eyes, as it may cause irritation.
What you need:
• SD LipidoCare Professional Analyzer
• SD LipidoCare Professional Lipid Profile Test Strips
• SD LipidoCare Professional Blood Glucose Test Strips
• The codechip
• SD Lipid Control Solution - Level 1 , Level 2
OR
SD Glucose Control Solution – Level M, Level H
- 40 -